Ammonia Analysis in High-Purity Hydrogen for Fuel Cell Vehicles Using an Agilent 8890 GC/8255 NCD System
Technical Reference Summary
Covered Models & Document Scope
Application note describing a GC/NCD method for trace ammonia analysis in hydrogen for fuel cell vehicles, using an Agilent 8890 GC with six-port gas sampling valve, 2 mL sample loop, NCD detector, and Agilent J&W Select Low Ammonia column (p/n CP8590). No product model variants or ordering codes are covered.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Sample loop volume | 2 mL p.3 ↗ | |
| Column | Select Low Ammonia (p/n CP8590) p.3 ↗ | |
| Burner temperature | 900 °C p.3 ↗ | |
| Burner pressure | 131 Torr p.3 ↗ | |
| Data rate | 5 Hz p.3 ↗ | |
| Calibration level L1 concentration | 37.1 ppb p.4 ↗ | |
| Calibration level L2 concentration | 49.5 ppb p.4 ↗ | |
| Calibration level L3 concentration | 79.2 ppb p.4 ↗ | |
| Calibration level L4 concentration | 99 ppb p.4 ↗ | |
| Calibration level L5 concentration | 198 ppb p.4 ↗ | |
| Calibration level L6 concentration | 495 ppb p.4 ↗ | |
| Response factor L1 | 0.228 p.4 ↗ | |
| Response factor L2 | 0.236 p.4 ↗ | |
| Response factor L3 | 0.218 p.4 ↗ | |
| Response factor L4 | 0.215 p.4 ↗ | |
| Response factor L5 | 0.175 p.4 ↗ | |
| Response factor L6 | 0.156 p.4 ↗ | |
| Linearity range | 37.1 to 495 ppb p.3 ↗ | |
| Quality control requirement | 100 ppb p.2 ↗ | ISO 14687-2019 / GB/T 37244-2018 |
Operating Limits
- Linearity range studied: 37.1 to 495 ppb p.3 ↗
- Quality control requirement (ISO 14687-2019, GB/T 37244-2018): 100 ppb p.2 ↗