Pesticide Analysis of Peppers Using Captiva EMR–GPF Cleanup and LC/MS/MS or GC/MS/MS

Technical Reference Summary

Covered Models & Document Scope

Application note describing a multiresidue pesticide determination method (240 pesticides) in mixed bell peppers using Agilent Captiva EMR–GPF passthrough cleanup followed by LC/MS/MS and GC/MS/MS; covers Captiva EMR–GPF cartridge 3 mL (part number 5610-2090) and compares it with Captiva EMR–HCF1, EMR–HCF2, EMR–GPD, EMR–LPD, U-dSPE with GCB, and vendor 1/vendor 2 dSPE cleanup.

Technical Data

Parameter Value Condition
LC/MS/MS injection volume 2 μL p.3 ↗ LC/MS/MS conditions
LC/MS/MS stop time 10 min p.3 ↗ LC/MS/MS conditions
LC/MS/MS gas temperature 120 °C p.3 ↗ LC/MS/MS conditions
LC/MS/MS gas flow 20 L/min p.3 ↗ LC/MS/MS conditions
LC/MS/MS nebulizer pressure 40 psi p.3 ↗ LC/MS/MS conditions
LC/MS/MS sheath gas heater temperature 225 °C p.3 ↗ LC/MS/MS conditions
LC/MS/MS low-pressure RF 60 V p.3 ↗ positive polarity
LC/MS/MS low-pressure RF 60 V p.3 ↗ negative polarity
LC/MS/MS polarity Positive and negative p.3 ↗ LC/MS/MS conditions
GC/MS/MS carrier gas Helium p.3 ↗ GC/MS/MS conditions
GC/MS/MS column 1 flow 1.0 mL/min p.3 ↗ GC/MS/MS conditions
GC/MS/MS column 2 flow 1.4 mL/min p.3 ↗ GC/MS/MS conditions
GC/MS/MS injection volume 1 µL p.3 ↗ cold splitless
GC/MS/MS inlet liner Agilent inlet liner, Ultra Inert, splitless p.3 ↗ GC/MS/MS conditions
Captiva EMR–GPF cartridge volume 3 mL p.3 ↗ Captiva EMR–GPF
Captiva EMR–GPF part number 5610-2090 p.3 ↗ Captiva EMR–GPF cartridge, 3 mL
Ceramic homogenizer pack size 100 pk p.3 ↗ 50 mL tubes, part number 5982-9313
Spiking level for recovery study 10 ng/g p.13 ↗ 230 targets
Recovery acceptance criterion 60 to 120 % p.16 ↗ accuracy
RSD acceptance criterion <20 % p.16 ↗ reproducibility
Matrix effect acceptance criterion (LC/MS/MS) 70 to 100 % p.16 ↗ LC/TQ only
Calibration curve acceptance criterion R2 >0.99 p.16 ↗ calibration curve
Number of pesticides targeted 240 p.1 ↗ multiresidue method
Number of LC-amenable pesticides 129 p.2 ↗ LC/MS/MS
Number of pesticides in quantitative analysis 230 p.10 ↗ bell pepper, LC/MS/MS and GC/MS/MS

Operating Limits

  • Recovery acceptance range: 60 to 120% p.16 ↗
  • RSD acceptance criterion: <20% p.16 ↗
  • Matrix effect acceptance range (LC/MS/MS): 70 to 100% p.16 ↗
  • Calibration curve acceptance criterion: R2 >0.99 p.16 ↗
  • Lowest spiking level used in recovery study: 10 ng/g p.13 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
Agilent Captiva EMR–GPF cartridge, 3 mL p.3 ↗ 5610-2090 Passthrough cleanup cartridge used in the method
Agilent ceramic homogenizers, 50 mL tubes, 100/pk p.3 ↗ 5982-9313 Sample preparation consumable