Application note covering characterization of multiple GLP-1 analogs (semaglutide, liraglutide, tirzepatide) using an Agilent 1290 Infinity II Bio LC System coupled to an Agilent 6545XT AdvanceBio LC/Q-TOF MS with ExD cell, controlled by MassHunter, ExDControl and ExDViewer software.

Technical Data

Parameter Value Condition
Column Agilent AdvanceBio Peptide Mapping 120Å, 2.1 × 150 mm, 2.7 µm p.3 ↗ LC method
Mobile Phase A LC/MS-grade water + 0.1% formic acid p.3 ↗ LC method
Mobile Phase B Acetonitrile + 0.1% formic acid p.3 ↗ LC method
Flow Rate 0.400 mL/min p.3 ↗ LC method
Injection Volume 10 µL p.3 ↗ LC method
Column Temperature 60 °C p.3 ↗ LC method
Ion Source Agilent Dual Jet Stream ESI source p.3 ↗ MS method
Polarity Positive p.3 ↗ MS method
Gas Temperature 325 °C p.3 ↗ MS method
Drying Gas Flow 10 L/min p.3 ↗ MS method
Nebulizer 35 psi p.3 ↗ MS method
Sheath Gas Temperature 275 °C p.3 ↗ MS method
Sheath Gas Flow 12 L/min p.3 ↗ MS method
Capillary Voltage 3,500 V p.3 ↗ MS method
Nozzle Voltage 2,000 V p.3 ↗ MS method
Fragmentor 175 V p.3 ↗ MS method
Skimmer 75 V p.3 ↗ MS method
Acquisition Rate 2 spectra/sec p.3 ↗ MS method
Acquisition Mode Extended Dynamic Range (2 GHz) p.3 ↗ MS method
Isolation Window Medium (4 amu) p.3 ↗ MS method
MS Spectrum Range m/z 120 to 3,200 p.3 ↗ MS method
MS/MS Spectrum Range m/z 120 to 3,200 p.3 ↗ MS method
Intensity Threshold (Abs) 2,000 p.3 ↗ Auto MS/MS
Intensity Threshold (Rel) 0.01% p.3 ↗ Auto MS/MS
Mass Error Tolerance 20 ppm p.3 ↗ Auto MS/MS

Operating Limits

  • MS spectrum range: m/z 120 to 3,200 p.3 ↗
  • MS/MS spectrum range: m/z 120 to 3,200 p.3 ↗
  • Column temperature: 60 °C p.3 ↗
  • Gas temperature: 325 °C p.3 ↗
  • Sheath gas temperature: 275 °C p.3 ↗