Dynamic MRM/Scan Mode, Sensitive Quantitation in Complex Foods by GC/TQ
Technical Reference Summary
Covered Models & Document Scope
Application note covering simultaneous dMRM/scan data acquisition on Agilent 8890/7000E and 8890/7010C triple quadrupole GC/MS (GC/TQ) systems for pesticide residue analysis in spinach, walnut and cayenne pepper matrices; no ordering codes or product variants are defined.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| GC system | Agilent 8890 with fast oven, auto injector p.3 ↗ | 8890/7000E and 8890/7010C |
| Mass spectrometer | 7000E / 7010C triple quadrupole p.2 ↗ | |
| Vacuum pump | Performance turbo p.3 ↗ | |
| Source temperature | 280 °C p.3 ↗ | |
| Acquisition mode | Simultaneous dMRM/scan p.3 ↗ | |
| Initial oven temperature | 60 °C p.3 ↗ | |
| Initial oven hold | 1 min p.3 ↗ | |
| Final temperature 1 | 170 °C p.3 ↗ | ramp 1 |
| Ramp rate 2 | 10 °C/min p.3 ↗ | |
| Final temperature 2 | 310 °C p.3 ↗ | ramp 2 |
| Final hold 2 | 2.25 min p.3 ↗ | ramp 2 |
| Total run time | 20 min p.3 ↗ | |
| Post run time | 1.5 min p.3 ↗ | |
| Equilibration time | 0.25 min p.3 ↗ | |
| Post run inlet temperature | 310 °C p.3 ↗ | |
| Post run total flow | 25 mL/min p.3 ↗ | |
| Column 1 | Agilent HP-5ms UI, 15 m × 0.25 mm, 0.25 µm p.3 ↗ | |
| He quench gas flow | 2.25 mL/min p.3 ↗ | |
| N2 collision gas flow | 1.5 mL/min p.3 ↗ | |
| Total MRMs (dMRM mode) | 614 p.3 ↗ | |
| Minimum dwell time | 6.85 ms p.3 ↗ | |
| Minimum cycle time | 69.8 ms p.3 ↗ | |
| Scan type | MS1 scan p.3 ↗ | full scan |
| Scan range | 45 to 450 m/z p.3 ↗ | full scan |
Operating Limits
- Calibration range: 0.1 to 5,000 ppb p.11 ↗
- Minimum dwell time: 6.85 ms p.3 ↗
- Minimum cycle time: 69.8 ms p.3 ↗