Analyse gasförmiger Produkte aus Technologien zur
Technical Reference Summary
Covered Models & Document Scope
Application Note (Agilent Technologies) describing a multidimensional GC configuration with TCD, FID and nickel catalyst for analysis of gaseous products from CO2 utilization technologies; covers H2, O2, N2, CH4, CO, CO2 and C1–C6 hydrocarbons; document is an application note, not a product datasheet, and lists no instrument model/size variants or ordering codes.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Detection range, lower limit | 0.1 ppm p.1 ↗ | overall system, TCD + FID with catalyst |
| Detection range, upper limit | 100 % p.1 ↗ | overall system, TCD + FID with catalyst |
| TCD detection limit | 100 ppm p.1 ↗ | H2, O2, N2, CH4, CO, CO2 |
| FID with catalyst detection limit | 0.1 ppm p.1 ↗ | CH4, CO, CO2, C2–C6 hydrocarbons |
| Analytes | H2, O2, N2, CH4, CO, CO2, C1–C6 hydrocarbons p.1 ↗ | |
| Detectors | TCD and FID p.1 ↗ | multidimensional configuration |
| Catalyst | Nickel catalyst p.1 ↗ | positioned at FID |
| Valve 1 switching | ON/OFF position p.3 ↗ | used to divert compounds with more than 2 carbon atoms (C2+) |
| Pressure controller | three-channel pressure controller p.3 ↗ |
Operating Limits
- Detection range of the method: 0.1 ppm to 100 % p.1 ↗
- TCD detection limit: 100 ppm p.1 ↗
- FID with catalyst detection limit: 0.1 ppm p.1 ↗