Induztr
Engineering Digest Document Viewer Related Models

Agilent LUMA Multichannel Vacuum Ultraviolet Detector on Agilent 8890 GC

Technical Reference Summary

Covered Models & Document Scope

Application note for the LUMA multichannel vacuum ultraviolet (VUV) detector used with an Agilent 8890 GC and J&W DB-1 column (30 m × 320 µm, 1 µm); covers detector evaluation with the Polar ISO Column Text Mix standard, 12-band configuration, and eight analytes at 7.8–250 ppm.

Technical Data

Parameter Value Condition
Injection volume 1 µL p.2 ↗
Inlet temperature 250 °C p.2 ↗ Split/Splitless, 5:1 split
Oven program initial temperature 40 °C p.2 ↗ hold for 1 min
Oven program ramp rate 20 °C/min p.2 ↗ to 260 °C
Oven program final temperature 260 °C p.2 ↗
Column flow 6.5 mL/min p.2 ↗ hydrogen carrier
LUMA temperature setpoint 275 °C p.2 ↗
Number of detector bands 12 p.2 ↗ simultaneously acquired
Standard concentration range 7.8 to 250 ppm p.3 ↗ six standards
Average peak area RSD < 2 % p.3 ↗ five consecutive injections, 7.8 and 250 ppm
Average R² 0.999 p.3 ↗ eight compounds
S/N for Aniline 415 p.3 ↗ 7.8 ppm, Band 7
S/N for 2-Chlorophenol 398 p.3 ↗ 7.8 ppm, Band 7
S/N for 1-Octanol 155 p.3 ↗ 7.8 ppm, Band 2
S/N for 2-Nonanone 169 p.3 ↗ 7.8 ppm, Band 2
S/N for 2-Dodecanol 152 p.3 ↗ 7.8 ppm, Band 2
S/N for Methyl Laurate (C12:0) 150 p.3 ↗ 7.8 ppm, Band 2
S/N for n-Heptadecane (C17) 170 p.3 ↗ 7.8 ppm, Band 2
S/N for n-Nonadecane (C19) 158 p.3 ↗ 7.8 ppm, Band 2
Peak symmetry for Aniline 0.95 p.3 ↗ Band 7
Peak symmetry for 2-Chlorophenol 0.98 p.3 ↗ Band 7
Peak symmetry for 1-Octanol 0.97 p.3 ↗ Band 2
Peak symmetry for 2-Nonanone 0.98 p.3 ↗ Band 2
Peak symmetry for 2-Dodecanol 0.94 p.3 ↗ Band 2
Peak symmetry for Methyl Laurate (C12:0) 1.00 p.3 ↗ Band 2

Operating Limits

  • LUMA detector temperature setpoint: 275 °C p.2 ↗
  • GC inlet temperature: 250 °C p.2 ↗
  • Oven temperature range: 40 °C to 260 °C p.2 ↗
  • Band 2 wavelength range: 130 to 143 nm p.3 ↗
  • Band 7 wavelength range: 189 to 193 nm p.3 ↗
  • Band 1 wavelength range: 118 to 130 nm p.3 ↗