Induztr
Engineering Digest Document Viewer Related Models

Analysis of Sulfur Compounds in High-Purity Hydrogen Using an Agilent GC/SCD

Technical Reference Summary

Covered Models & Document Scope

Application note covering analysis of sulfur compounds in hydrogen using an Agilent 8890 gas chromatograph configured with an Agilent 8355 sulfur chemiluminescence detector (SCD); no product model/size variants or ordering codes are given.

Technical Data

Parameter Value Condition
Sample introduction six-port gas sampling valve connected directly to an inert volatiles interface (VI) p.2 ↗ 8890 GC with 8355 SCD
Sample loop volume 2 mL p.2 ↗ 8890 GC with 8355 SCD
Split ratio 10:1 p.2 ↗ 8890 GC with 8355 SCD
Data acquisition rate 5 Hz p.3 ↗ 0.04 minutes
Calibration range 15 to 200 ppb p.5 ↗ sulfur compounds
Calibration level 15 ppb p.5 ↗ low level
Calibration mix concentration 0.992 µmol/mol, ppm p.3 ↗ Hydrogen sulfide (H2S)
Retention time 3.976 min p.5 ↗ Hydrogen sulfide
Correlation coefficient (R2) 0.9987 p.5 ↗ Hydrogen sulfide

Operating Limits

  • Calibration range: 15 to 200 ppb p.5 ↗
  • Data acquisition rate: 5 Hz / 0.04 minutes p.3 ↗
  • Sample loop volume: 2 mL p.2 ↗
  • Split ratio: 10:1 p.2 ↗