Analysis of Sulfur Compounds in High-Purity Hydrogen Using an Agilent GC/SCD
Technical Reference Summary
Covered Models & Document Scope
Application note covering analysis of sulfur compounds in hydrogen using an Agilent 8890 gas chromatograph configured with an Agilent 8355 sulfur chemiluminescence detector (SCD); no product model/size variants or ordering codes are given.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Sample introduction | six-port gas sampling valve connected directly to an inert volatiles interface (VI) p.2 ↗ | 8890 GC with 8355 SCD |
| Sample loop volume | 2 mL p.2 ↗ | 8890 GC with 8355 SCD |
| Split ratio | 10:1 p.2 ↗ | 8890 GC with 8355 SCD |
| Data acquisition rate | 5 Hz p.3 ↗ | 0.04 minutes |
| Calibration range | 15 to 200 ppb p.5 ↗ | sulfur compounds |
| Calibration level | 15 ppb p.5 ↗ | low level |
| Calibration mix concentration | 0.992 µmol/mol, ppm p.3 ↗ | Hydrogen sulfide (H2S) |
| Retention time | 3.976 min p.5 ↗ | Hydrogen sulfide |
| Correlation coefficient (R2) | 0.9987 p.5 ↗ | Hydrogen sulfide |
Operating Limits
- Calibration range: 15 to 200 ppb p.5 ↗
- Data acquisition rate: 5 Hz / 0.04 minutes p.3 ↗
- Sample loop volume: 2 mL p.2 ↗
- Split ratio: 10:1 p.2 ↗