Induztr
Engineering Digest Document Viewer Related Models

Elemental Analysis of Crude Oil and Residual Fuels using ASTM D8322

Technical Reference Summary

Covered Models & Document Scope

Application note covering analysis of crude oils and residual fuel oils per ASTM D8322-20 using three Agilent 4210 MP-AES instruments with Agilent 4107 Nitrogen Generator, in-house N2 supply, or a combination; elements V, Ni, Ca, Na, Al, Si, Zn, S in residual fuels and Fe, V, Ni, Ca, Na, K, S in crude oils.

Technical Data

Parameter Value Condition
Stabilization time 30 s p.2 ↗
Background correction Auto p.2 ↗
Air injection required Yes p.2 ↗
Nebulizer flow rate 0.5 L/min p.3 ↗ all elements
Iron wavelength 259.940 nm p.3 ↗ Fe
Vanadium wavelength 311.070 nm p.3 ↗ V
Nickel wavelength 341.476 nm p.3 ↗ Ni
Calcium wavelength 396.847 nm p.3 ↗ Ca
Sodium wavelength 588.995 nm p.3 ↗ Na
Aluminum wavelength 396.152 nm p.3 ↗ Al
Silicon wavelength 288.158 nm p.3 ↗ Si
Zinc wavelength 213.857 nm p.3 ↗ Zn
Potassium wavelength 766.491 nm p.3 ↗ K
Sulfur wavelength 181.972 nm p.3 ↗ S
Yttrium wavelength 371.029 nm p.3 ↗ Y (internal standard)
EGCM flow rate Medium p.3 ↗ Fe, Si, Zn, Y
EGCM flow rate High p.3 ↗ V, Ni, Ca, Na, Al, K
EGCM flow rate Low p.3 ↗ S