Structural Analysis of the Multimers of Bovine Serum Albumin by Ion Mobility Mass Spectrometry
Technical Reference Summary
Covered Models & Document Scope
Application note covering analysis of BSA multimers using the Agilent 6560 ion mobility LC/Q-TOF with an Agilent nanospray source (G1992A); solution conditions: 150 mM ammonium acetate, 0.1% formic acid, and 150 mM ammonium acetate (per text). No product model/size variants or ordering codes are given.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Instrument | 6560 ion mobility LC/Q-TOF p.1 ↗ | |
| Ionization source | Agilent nanospray source (G1992A) p.2 ↗ | |
| Sample | BSA (Sigma-Aldrich) p.2 ↗ | |
| Sample concentration | 20 mg/mL p.2 ↗ | |
| Flow rate | 266 nL/min p.2 ↗ | |
| Needle dimensions | 360 µm od, 75 µm id, 30 µm tip, 5 cm length p.2 ↗ | New Objective noncoated needle |
| Mobile phase 1 | 150 mM ammonium acetate p.2 ↗ | |
| Mobile phase 2 | 0.1% formic acid p.2 ↗ | |
| Drift tube length | ~80 cm p.2 ↗ | |
| High pressure funnel voltage | 200 V peak-to-peak p.3 ↗ | optimized for native BSA transmission |
| Trapping funnel voltage | 200 V peak-to-peak p.3 ↗ | optimized for native BSA transmission |
| Monomer m/z (charge 1) | 66,435 m/z p.4 ↗ | |
| Dimer m/z (charge 1) | 132,869 m/z p.4 ↗ | |
| Trimer m/z (charge 1) | 199,303 m/z p.4 ↗ | |
| Tetramer m/z (charge 1) | 265,737 m/z p.4 ↗ |