Induztr
Engineering Digest Document Viewer Related Models

ASTM D8351-22: Elemental Analysis of Biodiesel using MP-AES

Technical Reference Summary

Covered Models & Document Scope

Application study of biodiesel elemental analysis (Ca, Mg, P, K, Na) per ASTM D8351-22 using the Agilent 4210 MP-AES with organics kit (EGCM, OneNeb Series 2 nebulizer, double-pass glass cyclonic spray chamber), SPS 4 autosampler and optional Agilent 4107 nitrogen generator; results from five laboratories.

Technical Data

Parameter Value Condition
Read time 3 to 10 s p.2 ↗ 4210 MP-AES operating conditions
Number of replicates 3 p.2 ↗ 4210 MP-AES operating conditions
Stabilization time 30 s p.2 ↗ 4210 MP-AES operating conditions
Rinse time Approximately 60 s p.2 ↗ depends on lab
Nebulizer flow rate 0.5 p.3 ↗ all elements, organics kit
EGCM injection High p.3 ↗ Ca 396.847 nm
EGCM injection Medium p.3 ↗ Mg 285.213 nm
EGCM injection Medium p.3 ↗ P 213.618 nm
EGCM injection High p.3 ↗ K 766.491 nm
EGCM injection High p.3 ↗ Na 588.995 nm
Background correction Auto p.3 ↗ all elements
Wavelength 396.847 nm p.3 ↗ Calcium
Wavelength 285.213 nm p.3 ↗ Magnesium
Wavelength 213.618 nm p.3 ↗ Phosphorus
Wavelength 766.491 nm p.3 ↗ Potassium
Wavelength 588.995 nm p.3 ↗ Sodium
Method detection limit (MDL) 0.01 ppm p.4 ↗ Calcium 396.847 nm
Method detection limit (MDL) 0.02 ppm p.4 ↗ Magnesium 285.213 nm
Method detection limit (MDL) 0.19 ppm p.4 ↗ Phosphorus 213.618 nm
Method detection limit (MDL) 0.04 ppm p.4 ↗ Potassium 766.491 nm
Method detection limit (MDL) 0.02 ppm p.4 ↗ Sodium 588.995 nm

Operating Limits

  • Method detection limit, Calcium: 0.01 ppm p.4 ↗
  • Method detection limit, Magnesium: 0.02 ppm p.4 ↗
  • Method detection limit, Phosphorus: 0.19 ppm p.4 ↗
  • Method detection limit, Potassium: 0.04 ppm p.4 ↗
  • Method detection limit, Sodium: 0.02 ppm p.4 ↗