Analyse von Trinkwasser mit dem Agilent 8860 GC und dem 7697A
Technical Reference Summary
Covered Models & Document Scope
Agilent application note 5994-1239DEE (German, 15 October 2019): analysis of volatile halogenated and aromatic hydrocarbons and organophosphorus pesticides in drinking water using an Agilent 8860 GC with 7697A headspace sampler, FPD and Agilent DB-624 UI column (Best.-Nr. 122-1334UI); no instrument model/size variants or ordering codes are covered.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Headspace oven temperature | 100 °C p.3 ↗ | |
| Carrier gas | Stickstoff p.3 ↗ | |
| Vial purge flow | 60 ml/min p.3 ↗ | |
| Split flow | 50 ml/min p.3 ↗ | |
| Fuel gas air flow | 60 ml/min p.3 ↗ | FPD |
| Fuel gas H2 flow | 60 ml/min p.3 ↗ | FPD |
| Column | Agilent DB-624 UI (Best.-Nr. 122-1334UI) p.3 ↗ | Säule 1 |
| Standard concentration | 100 μg/ml p.4 ↗ | |
| Calibration levels | 200, 500 und 1000 µg/l p.4 ↗ | |
| Area reproducibility (RSD, n = 10) | 1,54 % p.5 ↗ | Chloroform |
| Area reproducibility (RSD, n = 6) | 1,36 % p.5 ↗ | Dichlorvos |
Operating Limits
- Minimum calibration coefficient R2 for all compounds: > 0,9995 p.5 ↗
- Maximum area RSD (n = 6): < 1,83 % p.5 ↗
- Maximum retention time RSD (n = 10): < 0,017 % p.5 ↗