A New AdvanceBio HIC Column for Drug to Antibody Ratio (DAR) Analysis of Antibody Drug Conjugates (ADC)

Technical Reference Summary

Covered Models & Document Scope

Application note for DAR analysis of cysteine-linked antibody-drug conjugates (ADCs) by HIC using the Agilent 1260 Infinity II bio-inert LC system (pump G5654A, multisampler G5668A with sample cooler option 100, multicolumn thermostat G7116A with bio-inert heat exchanger option 019, diode array detector WR G7115A with bio-inert flow cell option 028) with AdvanceBio HIC columns 4.6 × 100 mm (p/n 685975-908) and 4.6 × 30 mm (p/n 681975-908); covers DAR 0, DAR 2, DAR 4, DAR 6 and DAR 8 variants.

Technical Data

Parameter Value Condition
Column AdvanceBio HIC 4.6 × 100 mm mm p.2 ↗ p/n 685975-908
Mobile phase A 50 mM Sodium phosphate, pH 7.0 p.2 ↗
Mobile phase B 2 M Ammonium sulfate in 50 mM sodium phosphate, pH 7.0 p.2 ↗
Mobile phase C Propan-2-ol p.2 ↗
Mobile phase D HPLC grade water p.2 ↗
Flow rate 0.5 mL/min p.2 ↗
Column temperature 25 °C p.2 ↗
Injection volume 5 µL p.2 ↗
Total run time 31 minutes p.2 ↗
Gradient time 0 50 / 45 / 5 %A / %B / %C p.2 ↗ typical gradient profile
Gradient time 15 75 / 0 / 25 %A / %B / %C p.2 ↗ typical gradient profile
Gradient time 20 75 / 0 / 25 %A / %B / %C p.2 ↗ typical gradient profile
Gradient time 21 50 / 45 / 5 %A / %B / %C p.2 ↗ typical gradient profile
Gradient time 31 50 / 45 / 5 %A / %B / %C p.2 ↗ typical gradient profile
Software Agilent OpenLab 2.2 CDS p.2 ↗

Operating Limits

  • column temperature: 25 °C p.2 ↗
  • flow rate: 0.5 mL/min p.2 ↗
  • injection volume: 5 µL p.2 ↗
  • total run time: 31 minutes p.2 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
AdvanceBio HIC 4.6 × 100 mm p.2 ↗ 685975-908 HIC column
AdvanceBio HIC 4.6 × 30 mm p.2 ↗ 681975-908 HIC column