Osmotic Fragility of Red Blood Cells Quantified on the Agilent NovoCyte Flow Cytometer
Technical Reference Summary
Covered Models & Document Scope
Application note on flow cytometry osmotic fragility testing of red blood cells using the Agilent NovoCyte flow cytometer; covers sample preparation, DI water spiking, and effects of incubation time and temperature. No instrument model variants or ordering codes are described.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Blood dilution step 1 | 20 µL blood in 1 mL PBS p.3 ↗ | |
| Blood dilution step 2 | 10 µL diluted blood in 1.1 mL PBS p.3 ↗ | |
| Final PBS concentration | 55 % PBS p.3 ↗ | |
| Baseline sample volume | 20 µL p.3 ↗ | |
| Flow rate | 50 µL/minute p.3 ↗ | |
| DI water spike volume | 0.9 mL p.3 ↗ | |
| Additional acquisition volume | 100 µL p.3 ↗ | |
| Acquisition segment duration | 10 second p.3 ↗ | |
| Normal donor residual red cells | ~65 % RRC p.3 ↗ | |
| Residual red cells after overnight incubation | 38 % RRC p.3 ↗ | 4 °C |
| Sample analysis window after collection | within two hours p.3 ↗ | EDTA tubes |
Operating Limits
- Recommended test timing after blood collection: immediately after blood collection (within two hours) p.3 ↗
- Incubation temperature range tested: 4 to 37 °C p.3 ↗