High Sensitivity Determination of Lead in Soy Sauce
Technical Reference Summary
Covered Models & Document Scope
Application note for the high-sensitivity determination of lead in soy sauce using the Agilent 280Z Graphite Furnace Atomic Absorption Spectrometer (GFAAS) with Omega platform tube and PSD 120 autosampler; no instrument model/size variants or ordering codes are covered.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Instrument | 280Z AA p.1 ↗ | GFAAS |
| Background correction | Transverse Zeeman p.2 ↗ | |
| Autosampler | PSD 120 p.2 ↗ | |
| Graphite tube | Omega platform tube p.2 ↗ | |
| Furnace viewing camera | Tube-CAM p.2 ↗ | standard on 280Z AA |
| Software | Spectra 5.5 p.2 ↗ | SRM Wizard |
| Drying step 1 temperature | 95 °C p.3 ↗ | 50.0 s, Ar 0.3 L/min |
| Drying step 2 temperature | 120 °C p.3 ↗ | 20.0 s, Ar 0.3 L/min |
| Ashing step 3 temperature | 400 °C p.3 ↗ | 30.0 s, Ar 0.3 L/min |
| Ashing step 4 temperature | 500 °C p.3 ↗ | 20.0 s, Ar 0.3 L/min |
| Ashing step 5 temperature | 805 °C p.3 ↗ | 15.0 s, Ar 0.3 L/min |
| Ashing step 6 temperature | 805 °C p.3 ↗ | 10.0 s, Ar 0.3 L/min |
| Ashing step 7 temperature | 805 °C p.3 ↗ | 1.0 s, Ar 0.0 L/min |
| Atomization step 8 temperature | 2132 °C p.3 ↗ | 0.6 s, Ar 0.0 L/min, reading |
| Atomization step 9 temperature | 2132 °C p.3 ↗ | 2.0 s, Ar 0.0 L/min, reading |
| Cleanout step 10 temperature | 2500 °C p.3 ↗ | 2.0 s, Ar 0.3 L/min |
| Ash temperature (SRM) | 1000 °C p.3 ↗ | step 300 °C |
| Atomization temperature (SRM) | 2000 °C p.3 ↗ | step 300 °C |
| MDL in solution | 0.35 μg/kg p.4 ↗ | |
| MDL in sample | 18 μg/kg p.4 ↗ | |
| LOQ in sample | 59 μg/kg p.4 ↗ | |
| Maximum allowable limit (Pb in condiments) | 1000 μg/kg p.4 ↗ | GB 2762-2017 |
Operating Limits
- Maximum allowable Pb content in condiments (GB 2762-2017): ≤1.0 mg/kg p.4 ↗
- Maximum furnace program temperature: 2500 °C p.3 ↗