Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Technical Reference Summary

Covered Models & Document Scope

Application Note covering analysis of USP Method <467> residual solvents on the Agilent 8890 GC System with 7697A headspace sampler, dual FIDs, and J&W DB-Select 624 UI for 467 and J&W HP-INNOWax columns

Technical Data

Parameter Value Condition
Carrier gas Helium constant flow mode, 2 mL/min on column p.3 ↗
Inlet temperature 140 °C p.3 ↗
Oven temperature program 40 °C (hold 5 min) to 180 °C at 18 °C/min °C p.3 ↗
FID temperature (both channels) 250 °C p.3 ↗
Loop equilibration time 0.05 minutes p.3 ↗ 7697A headspace sampler
Software Agilent OpenLab CDS – Version 2.2 p.3 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
10-mL Clear crimp top headspace vials p.2 ↗ 5190–2285 Vials
Advanced Green nonstick inlet septa p.2 ↗ 5183–4759-100 Septa
Inert tee for capillary flow technology p.2 ↗ G3184-60065 Splitter
Short graphite ferrules for 0.1 to 0.32 mm columns, 10/pk p.2 ↗ 5080-8853 Ferrules
UltiMetal Plus flexible metal ferrules for 0.32 mm fused silica tubing, 10/pk p.2 ↗ G3188-27502 Ferrules
2 mm Splitless straight deactivated inlet liner p.2 ↗ 5181–8818 Inlet liner