Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | +5 °C above ambient temperature to 70 °C °C p.2 ↗ | BDS056-230V |
| Temperature uniformity at 37°C | 0.5 ± K K p.2 ↗ | BDS056-230V |
| Temperature fluctuation at 37°C | 0.3 ± K K p.2 ↗ | BDS056-230V |
| Heating up time to 37°C | 230 min p.2 ↗ | BDS056-230V |
| Recovery time after door was opened for 30 s at 37°C | 15 min p.2 ↗ | BDS056-230V |
| Rated Voltage | 230 V p.2 ↗ | BDS056-230V |
| Power frequency | 50/60 Hz p.2 ↗ | BDS056-230V |
| Nominal power | 0,3 kW p.2 ↗ | BDS056-230V |
| Unit fuse | 6,3 A p.2 ↗ | BDS056-230V |
| Phase (Nominal voltage) | 1~ p.2 ↗ | BDS056-230V |
| Interior volume | 62 L p.2 ↗ | BDS056-230V |
| Net weight of the unit (empty) | 35 kg p.2 ↗ | BDS056-230V |
| Load per rack | 15 kg p.2 ↗ | BDS056-230V |
| Permitted load | 45 kg p.2 ↗ | BDS056-230V |
| Wall clearance back | 160 mm p.2 ↗ | BDS056-230V |
| Wall clearance sidewise | 100 mm p.2 ↗ | BDS056-230V |
| Width net | 614 mm p.2 ↗ | BDS056-230V |
| Height net | 625 mm p.2 ↗ | BDS056-230V |
| Depth net | 577 mm p.2 ↗ | BDS056-230V |
| Interior width | 400 mm p.2 ↗ | BDS056-230V |
| Interior height | 440 mm p.2 ↗ | BDS056-230V |
| Interior depth | 350 mm p.3 ↗ | BDS056-230V |
| Energy consumption at 37°C | 25 Wh/h p.3 ↗ | BDS056-230V |
| Number of shelves (std./max.) | 1/3 p.3 ↗ | BDS056-230V |
| Rated Voltage | 120 V p.2 ↗ | BDS056UL-120V, 60 Hz |
Operating Limits
- Temperature range: +5 °C above ambient temperature to 70 °C p.2 ↗
- Wall clearance back: 160 mm p.2 ↗
- Wall clearance sidewise: 100 mm p.2 ↗
- Permitted load: 45 kg p.2 ↗
- Load per rack: 15 kg p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-3 p.1 ↗ | BDS | Product line: Solid.Line standard incubator |
| 4-6 p.1 ↗ | 056 | Model size (56 series, 62 L interior volume) |
| suffix p.1 ↗ | UL | UL variant for 120 V / 60 Hz supply with NEMA 5-15 plug (absent = standard 230 V variant) |
| suffix p.1 ↗ | 230V | Mains voltage variant 230 V 1~ 50/60 Hz, CEE 7/7 plug |
| suffix p.1 ↗ | 120V | Mains voltage variant 120 V 1~ 60 Hz, NEMA 5-15 plug |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| BDS056-230V p.1 ↗ | 9090-0016 | 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard | Standard version | |
| BDS056UL-120V p.1 ↗ | 9090-0017 | 120 V 1~ 60 Hz, NEMA 5-15 plug, Standard | UL version |