Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model FD S 56

Technical Reference Summary

Covered Models & Document Scope

Model FD-S 56 drying and heating chamber with forced convection, interior volume 55 L. Two standard variants: FDS056-230V (Art.-No. 9090-0018, 230 V 1~ 50/60 Hz, CEE 7/7 plug) and FDS056UL-120V (Art.-No. 9090-0019, 120 V 1~ 60 Hz, NEMA 5-15 plug).

Technical Data

Parameter Value Condition
Temperature range +10 °C above ambient temperature to 250 °C p.2 ↗ FDS056-230V
Temperature uniformity at 150°C 2.6 ± K p.2 ↗ FDS056-230V
Temperature fluctuation at 150°C 0.4 ± K p.2 ↗ FDS056-230V
Heating up time to 150°C 15 min p.2 ↗ FDS056-230V
Recovery time after door was opened for 30 s at 150°C 5 min p.2 ↗ FDS056-230V
Air change rate at 100°C 4 x/h p.2 ↗ FDS056-230V
Rated voltage 230 V p.2 ↗ FDS056-230V
Power frequency 50/60 Hz p.2 ↗ FDS056-230V
Nominal power 1,1 kW p.2 ↗ FDS056-230V
Unit fuse 6,3 A p.2 ↗ FDS056-230V
Phase (nominal voltage) 1~ p.2 ↗ FDS056-230V
Interior volume 55 L p.2 ↗ FDS056-230V
Net weight of the unit (empty) 35 kg p.2 ↗ FDS056-230V
Load per rack 15 kg p.2 ↗ FDS056-230V
Permitted load 45 kg p.2 ↗ FDS056-230V
Wall clearance back 160 mm p.2 ↗ FDS056-230V
Wall clearance sidewise 100 mm p.2 ↗ FDS056-230V
Width net 614 mm p.2 ↗ FDS056-230V
Height net 625 mm p.2 ↗ FDS056-230V
Depth net 577 mm p.2 ↗ FDS056-230V
Interior width 400 mm p.3 ↗ FDS056-230V
Interior height 440 mm p.3 ↗ FDS056-230V
Interior depth 310 mm p.3 ↗ FDS056-230V
Sound-pressure level 43 dB(A) p.3 ↗ FDS056-230V
Energy consumption at 150°C 290 Wh/h p.3 ↗ FDS056-230V

Operating Limits

  • Temperature range: +10 °C above ambient temperature to 250 °C p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗
  • Load per rack: 15 kg p.2 ↗
  • Permitted load: 45 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ FDS Product family: drying and heating chamber, Solid.Line
4-6 p.1 ↗ 056 Model size / interior volume class (55 L)
7-9 p.1 ↗ UL UL variant for 120 V / 60 Hz mains (absent for 230 V variant)
10-13 p.1 ↗ 230V / 120V Rated voltage variant

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
FDS056-230V p.1 ↗ 9090-0018 230 V 1~ 50/60 Hz, CEE 7/7 plug, standard version Standard model
FDS056UL-120V p.1 ↗ 9090-0019 120 V 1~ 60 Hz, NEMA 5-15 plug, standard version UL standard model

Related models