Model FP 115 drying and heating chamber with forced convection and program functions. Two standard variants: FP115-230V (Art.-No. 9010-0364, 230 V 1~ 50/60 Hz, CEE 7/7 plug) and FP115UL-120V (Art.-No. 9010-0368, 120 V 1~ 60 Hz, NEMA 5-20 plug).

Technical Data

Parameter Value Condition
Temperature range +10 °C above ambient temperature to 300 °C p.2 ↗ FP115-230V / FP115UL-120V
Temperature uniformity at 150°C 1.4 ± K p.2 ↗ FP115-230V / FP115UL-120V
Temperature fluctuation at 150°C 0.3 ± K p.2 ↗ FP115-230V / FP115UL-120V
Heating up time to 150°C 19 min p.2 ↗ FP115-230V / FP115UL-120V
Recovery time after door was opened for 30 s at 150°C 5 min p.2 ↗ FP115-230V / FP115UL-120V
Air change rate at 100°C 32 x/h p.2 ↗ FP115-230V / FP115UL-120V
Rated voltage 230 V p.2 ↗ FP115-230V
Rated voltage 120 V p.2 ↗ FP115UL-120V
Power frequency 50/60 Hz p.2 ↗ FP115-230V
Power frequency 60 Hz p.2 ↗ FP115UL-120V
Nominal power 1,4 kW p.2 ↗ FP115-230V / FP115UL-120V
Unit fuse 10,0 A p.2 ↗ FP115-230V
Unit fuse 16,0 A p.2 ↗ FP115UL-120V
Phase (nominal voltage) 1~ p.2 ↗ FP115-230V / FP115UL-120V
Interior volume 116 L p.2 ↗ FP115-230V / FP115UL-120V
Net weight of the unit (empty) 55 kg p.2 ↗ FP115-230V / FP115UL-120V
Load per rack 30 kg p.2 ↗ FP115-230V / FP115UL-120V
Permitted load 150 kg p.2 ↗ FP115-230V / FP115UL-120V
Wall clearance back 160 mm p.2 ↗ FP115-230V / FP115UL-120V
Wall clearance sidewise 100 mm p.2 ↗ FP115-230V / FP115UL-120V
Width net 714 mm p.2 ↗ FP115-230V / FP115UL-120V
Height net 732 mm p.2 ↗ FP115-230V / FP115UL-120V
Depth net 636 mm p.2 ↗ FP115-230V / FP115UL-120V
Sound-pressure level 43 dB(A) p.3 ↗ FP115-230V / FP115UL-120V
Number of shelves (std./max.) 2/5 p.3 ↗ FP115-230V / FP115UL-120V

Operating Limits

  • Temperature range: +10 °C above ambient temperature to 300 °C p.2 ↗
  • Permitted load: 150 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ FP Series: drying and heating chamber with forced convection and program functions
3-5 p.1 ↗ 115 Model size (interior volume class, 116 L)
suffix p.1 ↗ UL UL variant for 120 V / 60 Hz supply with NEMA 5-20 plug (absent = standard 230 V variant)
voltage field p.1 ↗ 230V Rated voltage 230 V, 1~, 50/60 Hz, CEE 7/7 plug, unit fuse 10,0 A
voltage field p.1 ↗ 120V Rated voltage 120 V, 1~, 60 Hz, NEMA 5-20 plug, unit fuse 16,0 A

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
FP115-230V p.1 ↗ 9010-0364 230 V 1~ 50/60 Hz, 10,0 A, CEE 7/7 plug, Standard Standard model version
FP115UL-120V p.1 ↗ 9010-0368 120 V 1~ 60 Hz, 16,0 A, NEMA 5-20 plug, Standard UL model version