Data Sheet Model MKFT 720
Technical Reference Summary
Covered Models & Document Scope
Model MKFT 720 dynamic climate chamber, 734 L interior volume. Two variants: MKFT720-400V (400 V 3~/N/PE 50 Hz, standard, Art.-No. 9020-0384) and MKFT720-480V-C (480 V 3~ 60 Hz, with voltage and frequency converter, Art.-No. 9020-0360).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | -70...180 °C p.3 ↗ | MKFT720-400V |
| Temperature range | -70...180 °C p.3 ↗ | MKFT720-480V-C |
| Temperature uniformity dependent on set | 0.3...2.0 ± K p.3 ↗ | MKFT720-400V |
| Temperature fluctuation dependent on set | 0.1...0.5 ± K p.3 ↗ | MKFT720-400V |
| Cooling down time from 180°C to -70°C | 120 min p.3 ↗ | MKFT720-400V |
| Temperature range with humidity | 10...95 °C p.3 ↗ | MKFT720-400V |
| Humidity range | 10...98 % RH p.3 ↗ | MKFT720-400V |
| Max. heat compensation at 25°C and 90% RH | 800 W p.3 ↗ | MKFT720-400V |
| Rated voltage | 400 V p.3 ↗ | MKFT720-400V |
| Rated voltage | 480 V p.3 ↗ | MKFT720-480V-C |
| Power frequency | 50 Hz p.3 ↗ | MKFT720-400V |
| Power frequency | 60 Hz p.3 ↗ | MKFT720-480V-C |
| Nominal power | 11 kW p.3 ↗ | MKFT720-400V |
| Unit fuse | 32,0 A p.3 ↗ | MKFT720-400V |
| Phase (nominal voltage) | 3~/N/PE p.3 ↗ | MKFT720-400V |
| Phase (nominal voltage) | 3~ p.3 ↗ | MKFT720-480V-C |
| Interior volume | 734 L p.3 ↗ | MKFT720-400V |
| Net weight of the unit (empty) | 635 kg p.3 ↗ | MKFT720-400V |
| Net weight of the unit (empty) | 887 kg p.3 ↗ | MKFT720-480V-C |
| Load per rack | 40 kg p.3 ↗ | MKFT720-400V |
| Permitted load | 160 kg p.3 ↗ | MKFT720-400V |
| Sound-pressure level | 69 dB(A) p.4 ↗ | MKFT720-400V |
| Sound-pressure level | 72 dB(A) p.4 ↗ | MKFT720-480V-C |
| Energy consumption at 25°C and 60% RH | 2,200 Wh/h p.4 ↗ | MKFT720-400V |
| Number of shelves (std./max.) | 1/11 p.4 ↗ | MKFT720-400V |
Operating Limits
- Temperature range: -70...180 °C p.3 ↗
- Temperature range with humidity: 10...95 °C p.3 ↗
- Humidity range: 10...98 % RH p.3 ↗
- Wall clearance back: 300 mm p.4 ↗
- Wall clearance sidewise: 200 mm p.4 ↗
- Permitted load: 160 kg p.3 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-5 p.1 ↗ | MKFT7 | Series MKFT, size 720 |
| suffix p.1 ↗ | 400V | Standard variant, 400 V 3~/N/PE 50 Hz |
| suffix p.2 ↗ | 480V-C | Variant with voltage and frequency converter, 480 V 3~ 60 Hz |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| MKFT720-400V p.1 ↗ | 9020-0384 | 400 V 3~/N/PE 50 Hz, standard | Standard model | |
| MKFT720-480V-C p.2 ↗ | 9020-0360 | 480 V 3~ 60 Hz, with voltage and frequency converter | Converter variant |