Data Sheet Model VDL 23
Technical Reference Summary
Covered Models & Document Scope
BINDER VDL 23 vacuum drying chamber for flammable solvents, 24 L interior volume. Two voltage variants: VDL023-230V (Art.-No. 9630-0009, 200...230 V 1~ 50/60 Hz, CEE 7/7 plug) and VDL023-120V (Art.-No. 9630-0013, 100...120 V 1~ 50/60 Hz, NEMA 5-15 plug). ATEX classification EX II 2/3/- G IIB T3 Gb/Gc/- X.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | +9 °C above ambient temperature to 110 °C p.2 ↗ | VDL 23 |
| Temperature uniformity at 100°C | 1.0 ± K p.2 ↗ | VDL 23 |
| Temperature fluctuation at 100°C | 0.1 ± K p.2 ↗ | VDL 23 |
| Heating up time to 100°C | 110 min p.2 ↗ | VDL 23 |
| Leak rate | 0,01 bar/h p.2 ↗ | VDL 23 |
| Pressure fluctuation over time at 50 mbar | 1 ± mbar p.2 ↗ | VDL 23 |
| Pressure fluctuation over time at 400 mbar | 70 ± mbar p.2 ↗ | VDL 23 |
| Rated voltage | 200...230 V p.2 ↗ | VDL023-230V |
| Rated voltage | 100...120 V p.2 ↗ | VDL023-120V |
| Power frequency | 50/60 Hz p.2 ↗ | VDL 23 |
| Nominal power | 0,9 kW p.2 ↗ | VDL 23 |
| Unit fuse | 6,3 A p.2 ↗ | VDL023-230V |
| Unit fuse | 10,0 A p.2 ↗ | VDL023-120V |
| Phase (nominal voltage) | 1~ p.2 ↗ | VDL 23 |
| Vacuum connection with small flange | 16 DN mm p.2 ↗ | VDL 23 |
| Measuring access port with small flange | 16 DN mm p.2 ↗ | VDL 23 |
| Interior volume | 24 L p.2 ↗ | VDL 23 |
| Net weight of the unit (empty) | 72 kg p.2 ↗ | VDL 23 |
| Load per rack | 20 kg p.2 ↗ | VDL 23 |
| Permitted load | 50 kg p.2 ↗ | VDL 23 |
| Wall clearance back | 100 mm p.2 ↗ | VDL 23 |
| Wall clearance sidewise | 70 mm p.2 ↗ | VDL 23 |
| Width net | 523 mm p.3 ↗ | VDL 23 |
| Height net | 698 mm p.3 ↗ | VDL 23 |
| Depth net | 413 mm p.3 ↗ | VDL 23 |
Operating Limits
- Temperature range: +9 °C above ambient temperature to 110 °C p.2 ↗
- Permitted load: 50 kg p.2 ↗
- Load per rack: 20 kg p.2 ↗
- Wall clearance back: 100 mm p.2 ↗
- Wall clearance sidewise: 70 mm p.2 ↗
- ATEX classification: EX II 2/3/- G IIB T3 Gb/Gc/- X p.1 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-6 p.1 ↗ | VDL023 | Model series VDL, size 23 |
| field 7 p.1 ↗ | - | Separator |
| field 8-10 p.1 ↗ | 120 / 230 | Rated voltage variant: 120 V (100...120 V) or 230 V (200...230 V) |
| field 11 p.1 ↗ | V | Voltage unit |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| VDL023-230V p.1 ↗ | 9630-0009 | 200...230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard | 230 V variant | |
| VDL023-120V p.1 ↗ | 9630-0013 | 100...120 V 1~ 50/60 Hz, NEMA 5-15 plug, Standard | 120 V variant |