Cary Eclipse Fluorescence Spectrophotometer User's Guide
Technical Reference Summary
Covered Models & Document Scope
Covers the Agilent Cary Eclipse Fluorescence Spectrophotometer system, including the standard single cell holder, optional sample holders, nitrogen purging accessories, Cary Eclipse WinFLR software, and the National Instruments USB-GPIB-HS converter. Manual part number 8510175800, Issue 7, September 2016.
Section-by-Section Technical Breakdown
Defines environmental conditions for transport and operation, instrument weight and dimensions, power requirements, electrical connections, fuse rating, nitrogen supply pressure, and computer placement. The instrument is for indoor use only, Pollution degree 2, Installation Category II (EN 61010-1).
- Operating temperature within performance specifications: 10–35 °C at 0–853 m altitude; 10–25 °C at 853–2133 m altitude.
- Recommended ambient temperature for optimum analytical performance: 20–25 °C, held constant to within ±2 °C throughout the working day.
- Power: 100–240 VAC, 50–60 Hz, approximately 180 VA.
- Fuse: T3.15 AH 250 V (5 x 20 mm) IEC127 Sheet 5.
- Nitrogen supply: 70 kPa (10 psi).
- Front accessory connection: 25-pin D-range connector; rear: IEEE 488 (GPIB Cary Eclipse system connection).
- Sample compartment connectors: 9-pin, 15-pin, and 25-pin D-range connectors (low voltage AC/DC).
Describes installation of the standard single cell holder, cuvette height and pitch adjustment, and alignment of the cell holder using the Align application and a fluorescent sample. Includes mechanical adjustment details and required tools.
- Cell holder base is positioned on locating pins and locked by turning the lever to the right.
- Cuvette height is raised by turning the cell lifter clockwise and lowered counter-clockwise; downward movement requires pushing the cuvette down due to spring tension.
- Pitch adjustment screw requires a 2.5 mm hex key; clockwise raises the front of the cell holder.
- Alignment uses a fluorescent sample (e.g., anthracene) in a microcell and a 540 nm wavelength to check beam passage.
- Pitch adjustment is optimized by observing the intensity of light striking white paper in the beam path.
Specifies requirements for the nitrogen purging system, including gas quality, pressure regulation, supply tubing, gas manifold, and flowmeter. The purging system is not supplied by Agilent.
- Liquid nitrogen (with heat exchanger) is recommended; compressed nitrogen must be dry, oil-free, and uncontaminated.
- Never use compressed nitrogen from a supplier using oil or water in the compression process.
- Operating pressure is from 70 kPa (10 psi); use an appropriate regulator and gauge.
- Supply tubing: clean, flexible PVC or equivalent, 6 mm (1/4 in) internal diameter; never use rubber tubing.
- Flowmeter should monitor flow rates between 0 and 30 liters per minute (64 cubic feet per hour).
Provides step-by-step instructions for installing, registering, and uninstalling the Cary Eclipse WinFLR software and Help, and for installing the National Instruments USB-GPIB-HS converter. Includes notes on installation time and driver prompts.
- Install Cary Eclipse WinFLR software before installing the USB-GPIB-HS converter.
- During .Net Framework and GPIB driver installation, the computer may appear frozen for 3 to 5 minutes; do not exit the installation.
- The Product Key is found on the cover of the Agilent Cary Eclipse software CD/DVD case.
- During uninstallation, click 'No' when prompted to remove drivers/files that other programs may rely on.
- Read the Software Status Bulletin and Agilent Cary Eclipse WinFLR Release Notes.pdf in the install directory.
Details the procedure for removing and installing the xenon flash lamp module, including safety precautions, and the alignment procedure using the Align application with specified instrument parameters. The lamp module must be aligned after replacement.
- Always disconnect the Cary Eclipse from the mains supply before changing the lamp module.
- Remove the lamp access panel, turn the instrument onto its right side, remove four thumb screws from the lamp module cover, and loosen three bolts marked with a black dot.
- Disengage two guide pins at the rear and disconnect the plug by squeezing the clip.
- Alignment parameters: Ex. Wavelength 540 nm, Ex. Slit 1.5 nm, Ave. time 0.1 s, PMT voltage Medium.
- Adjust the bottom left screw (vertical) and top right screw (horizontal) for maximum signal using the supplied 5.5 mm nut driver.
- Click 'Save conditions' to store instrument parameters and the maximum reading.
Describes the location and replacement of the two fuses at the rear of the instrument, fuse code interpretation, and the checking procedure. Repeated fuse blowing may indicate other instrument problems.
- The Cary Eclipse contains two fuses located at the back of the instrument.
- Replace a blown fuse with one of the same type and rating as indicated in Chapter 2 (T3.15 AH 250 V).
- Fuse code: 'T' = time lag, 'F' = fast-acting; current rating in amperes; 'H' = heavy, 'L' = low breaking capacity; voltage rating in volts.
- Always ensure the code on the fuse cap matches the information printed next to the fuse holders.
- If a fuse repeatedly blows, it may indicate other problems with the instrument.
Provides solutions for common problems: instrument offline, Connect button instead of Start, and instrument performance testing not meeting specifications. Includes checks for software restart, cable connections, initialization, lamp pulsing, and lamp alignment.
- Instrument offline: restart computer after software installation, check USB-GPIB-HS cable connection, ensure initialization tests complete before starting software.
- Connect button instead of Start: check for other Cary applications running; click Connect to gain control if they are not collecting data.
- Performance test failure: ensure sample compartment is empty, cable retaining screws are tightened, lamp is pulsing (green power indicator flashing), and lamp is correctly aligned.
- If the lamp is not pulsing, there may be a hardware conflict.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Operating temperature within performance specifications is 10–35 °C at altitudes 0–853 m and 10–25 °C at altitudes 853–2133 m. p.18 ↗ | Install the instrument in a temperature-controlled laboratory; higher altitudes require reduced ambient temperature to maintain performance. | Whole series |
| For optimum analytical performance, ambient temperature should be 20–25 °C and held constant to within ±2 °C throughout the entire working day. p.18 ↗ | Use a stable laboratory HVAC system to avoid drift in fluorescence measurements. | Whole series |
| Power requirement: 100–240 VAC, 50–60 Hz, approximately 180 VA. p.19 ↗ | Ensure the mains supply and circuit capacity are adequate; use the country-specific Agilent-supplied power cord. | Whole series |
| Fuse rating: T3.15 AH 250 V (5 x 20 mm) IEC127 Sheet 5. p.20 ↗ | Replace blown fuses only with the specified type and rating; verify the code on the fuse cap matches the label next to the fuse holder. | Whole series |
| Nitrogen supply operating pressure is from 70 kPa (10 psi). p.20 ↗ | Use an appropriate pressure regulator and gauge to maintain consistent nitrogen pressure for purging. | Whole series with nitrogen purging |
| Nitrogen supply tubing must be clean, flexible PVC or equivalent with 6 mm (1/4 in) internal diameter; never use rubber tubing. p.25 ↗ | Rubber tubing may introduce talc into the optical system, causing contamination. | Whole series with nitrogen purging |
| Flowmeter should be capable of monitoring flow rates between 0 and 30 liters per minute (64 cubic feet per hour). p.26 ↗ | Select a flowmeter with adequate range for the purging system. | Whole series with nitrogen purging |
| The lamp module contains a high voltage xenon flashlamp operating at 1260 kPa. p.8 ↗ | Never disassemble the lamp module or look down the hole; risk of electric shock and eye damage. | Whole series |
| Lamp alignment parameters: Ex. Wavelength 540 nm, Ex. Slit 1.5 nm, Ave. time 0.1 s, PMT voltage Medium. p.35 ↗ | Use these settings in the Align application when aligning the lamp module after replacement. | Whole series |
| Lamp alignment requires adjusting the bottom left screw (vertical) and top right screw (horizontal) for maximum signal using the supplied 5.5 mm nut driver. p.35 ↗ | Proper alignment is essential for optimum instrument performance after lamp module replacement. | Whole series |
| The single cell holder pitch adjustment screw requires a 2.5 mm hex key. p.24 ↗ | Keep a 2.5 mm hex key available for cell holder alignment. | Whole series with single cell holder |
| During software installation, the .Net Framework and GPIB driver installation may take 3 to 5 minutes and the computer may appear frozen. p.27 ↗ | Do not exit the installation during this period; wait for completion. | Cary Eclipse WinFLR software |
| The instrument weighs over 30 kg (66 lb); use two or more people when lifting or carrying. p.21 ↗ | Plan for safe handling during installation or relocation to avoid injury or equipment damage. | Whole series |
| The Cary Eclipse is classified Equipment Class I and requires an integral ground connection between the metal base and the three-pin earth-grounded receptacle. p.9 ↗ | Ensure proper grounding to avoid electric shock hazard. | Whole series |
| The instrument is suitable for indoor use only, Pollution degree 2, Installation Category II (EN 61010-1). p.18 ↗ | Do not install in outdoor or harsh environments; ensure the electrical installation meets Category II requirements. | Whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Instrument offline p.42 ↗ | Computer not restarted after software installation; main instrument cable (USB-GPIB-HS) not properly connected; instrument initialization not completed before starting software. | Restart the computer after installing Cary Eclipse WinFLR software; check the connection of the USB-GPIB-HS cable; ensure the instrument has completed initialization tests before starting the software; contact local Agilent office if problem persists. | |
| Connect button instead of Start p.42 ↗ | Another Cary application is running and may have control of the instrument. | Check for other Cary applications running; if they are not collecting data, click the 'Connect' button to get control of the instrument. | |
| Instrument performance tests do not meet specifications p.43 ↗ | Sample compartment not empty; cable connecting instrument to computer not correctly connected or retaining screws not tightened; lamp not pulsing during initialization (indicated by green power indicator not flashing); lamp not correctly aligned; possible hardware conflict. | Ensure sample compartment is empty; check cable connection and tighten retaining screws; verify lamp is pulsing (green power indicator flashing, audible lamp and monochromator/filter wheel movement); align the lamp module as described in Chapter 4; if lamp is not pulsing, check for hardware conflict. | Chapter 4 for lamp alignment |
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Non-operating (transport) altitude | 0–2133 m p.18 ↗ | Non-operating (transport) |
| Non-operating (transport) altitude | 0–7000 ft p.18 ↗ | Non-operating (transport) |
| Non-operating (transport) temperature | 5–45 °C p.18 ↗ | Non-operating (transport) |
| Non-operating (transport) temperature | 41–113 °F p.18 ↗ | Non-operating (transport) |
| Non-operating (transport) relative humidity, non-condensing | 20–80 % p.18 ↗ | Non-operating (transport) |
| Operating altitude within performance specifications | 0–853 m p.18 ↗ | Operating within performance specifications |
| Operating altitude within performance specifications | 0–2800 ft p.18 ↗ | Operating within performance specifications |
| Operating altitude within performance specifications | 853–2133 m p.18 ↗ | Operating within performance specifications |
| Operating altitude within performance specifications | 2800–7000 ft p.18 ↗ | Operating within performance specifications |
| Operating temperature within performance specifications | 10–35 °C p.18 ↗ | Operating within performance specifications, 0–853 m |
| Operating temperature within performance specifications | 50–95 °F p.18 ↗ | Operating within performance specifications, 0–853 m |
| Operating temperature within performance specifications | 10–25 °C p.18 ↗ | Operating within performance specifications, 853–2133 m |
| Operating temperature within performance specifications | 50–77 °F p.18 ↗ | Operating within performance specifications, 853–2133 m |
| Operating relative humidity, non-condensing | 8–80 % p.18 ↗ | Operating within performance specifications |
| Weight, packed | 54 kg p.19 ↗ | Packed |
| Weight, packed | 119 lb p.19 ↗ | Packed |
| Weight, unpacked | 31 kg p.19 ↗ | Unpacked |
| Weight, unpacked | 68 lb p.19 ↗ | Unpacked |
| Dimensions (width x depth x height), packed | 820 x 760 x 560 mm p.19 ↗ | Packed |
| Dimensions (width x depth x height), packed | 32 x 30 x 22 in p.19 ↗ | Packed |
| Dimensions (width x depth x height), unpacked | 600 x 610 x 280 mm p.19 ↗ | Unpacked |
| Dimensions (width x depth x height), unpacked | 24 x 24 x 11 in p.19 ↗ | Unpacked |
| Voltage | 100–240 VAC p.19 ↗ | |
| Frequency | 50–60 Hz p.19 ↗ | |
| Consumption | 180 VA p.19 ↗ | approximately |
| Mains inlet coupler | 3/2 A 120/250 VAC 50–60 Hz IEC type p.19 ↗ | |
| Fuse | T3.15 AH 250 V (5 x 20 mm) IEC127 Sheet 5 p.20 ↗ | |
| Nitrogen supply pressure | 70 kPa p.20 ↗ | |
| Nitrogen supply pressure | 10 psi p.20 ↗ |