Operating Instructions — vacuum pump (BA 150-3-EN, ed. 2.0 26.07.2023, EN)
Technical Reference Summary
Covered Models & Document Scope
Covers the oil-flooded rotary vane vacuum pumps V-VCA 25 and V-VCE 25 (V Series), including the VCE 25 (09) variant with selectable fine/rough vacuum operating range. Document edition 2.0, 26.07.2023, BA 150-3-EN.
Section-by-Section Technical Breakdown
Describes the construction of the single-stage oil-flooded rotary vane pumps V-VCA 25 and V-VCE 25, including vacuum connection, exhaust air outlet, filter housing, cooling air paths, oil filling point, oil sight glass, oil discharge point, gas ballast valve and oil separator housing. The two types are identical in construction and differ only in achievable vacuum.
- Single-stage oil-flooded rotary vane design; oil seals gaps, lubricates and dissipates compression heat.
- Integrated mesh filter on suction side; oil and oil mist separator on outlet side returns oil to circulation.
- Fan between pump housing and motor provides intensive air cooling; fan housing protects against contact.
- Integrated non-return valve prevents ventilation of the evacuated system after shutdown and prevents oil hammering on restart.
- Standard gas ballast valve prevents condensation of small amounts of water vapour; reinforced gas ballast available for higher vapour loads (Info I 200).
- Driven via coupling by a flanged three-phase standard motor.
- Hot surfaces > 70 °C are marked.
Defines the permissible continuous intake pressure ranges for V-VCA and V-VCE at 50 Hz and 60 Hz, and gives limits for evacuating closed systems and for frequent switching.
- Nominal pumping capacity 25 m3/h at 50 Hz with free intake.
- V-VCA 50 Hz: 0.5 to 130 mbar (abs.); 60 Hz: 0.5 to 100 mbar (abs.).
- V-VCE 50 Hz: 10 to 500 mbar (abs.); 60 Hz: 10 to 400 mbar (abs.).
- Continuous operation outside these ranges risks oil leaking through the outlet opening.
- When evacuating closed systems, volume to be evacuated must be no more than 2 % of nominal pumping capacity.
- Frequent switching (approx. 10 times per hour) or high ambient/inlet temperatures may exceed motor winding and bearing temperature limits; contact manufacturer.
- Outdoor installation requires protection from environmental influences (e.g. protective roof).
Covers preparation of installation, mechanical erection, pipe connection and motor connection. Specifies clearance, leveling, cooling-air and back-pressure requirements.
- Machine must be freely accessible from all sides; ventilation grids and holes must not be closed.
- Provide at least 40 cm space around the machine for maintenance.
- Machine may only be operated when set up horizontally; max. inclination 1° in all directions.
- Secure machine against tipping over and falling; floor must be plane and even and able to carry the machine weight.
- Cooling air inlets and outlets must be at least 20 cm from adjacent walls; exhaust cooling air must not be re-sucked.
- Ambient temperature min. +5 °C, max. +40 °C.
- Exhaust air opening must not be closed or restricted; no filter may be installed on the exhaust air side.
- Counter pressures on exhaust side only permissible up to +100 mbar.
- Only screw in pipes by hand; use flexible connections if necessary.
- Electrical installation only by qualified electrician observing EN 60204; main switch and motor protection switch installed by operating company.
- Motors comply with DIN EN 60034, protection class IP 55, insulation class F.
- Unless otherwise stated on motor data plate: ±5 % voltage deviation and ±2 % frequency deviation permitted.
Describes installation checks, rotation direction check, setting the operating range on VCE 25 (09), daily oil level check, and decommissioning/storage/recommissioning.
- Installation check includes no transport/assembly damage, safe horizontal standing, correct and leak-tight pipe connections, tight screw/flange connections, electrical installation per connection diagram, adequate room ventilation, oil filled and level checked, pump and pipelines cleaned.
- Check rotation direction with phase sequence indicator (anti-clockwise rotating field); keep 1 m distance from outlet and inlet connections.
- Start motor briefly (max. two seconds); looking at motor fan, it must rotate clockwise.
- After correcting rotation direction, run motor and stop after 2 minutes to top oil up to upper edge of sight glass; repeat until all oil pipes are filled.
- Filling point must not be open when pump is running.
- For VCE 25 (09), fine vacuum or rough vacuum operating range is selected via adjustment bolt in pipe under filter housing.
- Check oil level every day; if leaks occur or exhaust air contains visible oil mist, determine cause immediately.
Covers maintenance table, cleaning of vacuum pump and air filter, lubricating oil refilling and oil change, oil separator insert, motor and coupling, repair/service and spare parts.
- Oil viscosity must comply with ISO VG 100 as per DIN ISO 3448.
- Clean ventilation slots and mesh filter; clean filter disc of gas ballast valve.
- Change oil separator insert when dirty.
- Replace coupling rubbers when worn.
- Repairs only by authorised workshops using original spare parts.
Contains the table of malfunctions with symptoms, causes, elimination actions and reference chapters. Covers motor protection switch trips, insufficient pumping capacity, failure to reach final pressure, overheating, visible oil mist, strange noises and water in lubricating oil.
- Never put the pump back into operation after automatic shut-off without unequivocally finding and remedying the reason.
- Malfunction table references chapters 2.4, 5.1, 5.3, 5.5, 7.2, 7.4, 7.5, 7.6, 7.7, 7.8.2 and 7.10.
- Contact Elmo Rietschle Service for other malfunctions or those that cannot be eliminated.
Technical data chapter referenced for weight, sound pressure level and other performance data; detailed values are in data sheet D 150-3.
- Bearing surface must be designed to carry the machine weight (see chapter 10).
- Measured sound pressure level is given in chapter 10.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Nominal pumping capacity is 25 m3/h at 50 Hz with free intake. p.16 ↗ | Use as the base sizing value for 50 Hz systems; refer to data sheet D 150-3 for pumping capacity vs. intake pressure. | V-VCA 25 and V-VCE 25 |
| V-VCA continuous intake pressure range: 0.5 to 130 mbar (abs.) at 50 Hz and 0.5 to 100 mbar (abs.) at 60 Hz. p.16 ↗ | Select V-VCA for fine vacuum duty; continuous operation outside this range risks oil leakage through the outlet. | V-VCA 25 |
| V-VCE continuous intake pressure range: 10 to 500 mbar (abs.) at 50 Hz and 10 to 400 mbar (abs.) at 60 Hz. p.16 ↗ | Select V-VCE for rough vacuum duty; continuous operation outside this range risks oil leakage through the outlet. | V-VCE 25 |
| When evacuating closed systems, the volume to be evacuated must be no more than 2 % of the nominal pumping capacity. p.16 ↗ | Size the evacuated volume accordingly to avoid operating outside the intended use. | V-VCA 25 and V-VCE 25 |
| Maximum permissible back pressure on the exhaust side is +100 mbar. p.8 ↗ | Exhaust lines must be sized and routed to stay below this limit; no filter may be installed on the exhaust air side. | V-VCA 25 and V-VCE 25 |
| Permitted ambient and inlet temperature range is 5 to 40 °C. p.8 ↗ | Install in a ventilated room within this range; contact manufacturer for temperatures outside this range. | V-VCA 25 and V-VCE 25 |
| Machine may only be operated when set up horizontally with max. inclination 1° in all directions. p.17 ↗ | Mount on a plane, even floor able to carry the machine weight; secure against tipping and falling. | V-VCA 25 and V-VCE 25 |
| Cooling air inlets and outlets must be at least 20 cm away from adjacent walls. p.17 ↗ | Maintain clearance to prevent recirculation of hot cooling air and overheating. | V-VCA 25 and V-VCE 25 |
| Provide at least 40 cm space around the machine for maintenance work. p.17 ↗ | Layout the installation to allow access to filter housing, oil filling point, oil sight glass, oil drain, gas ballast and oil separator housing. | V-VCA 25 and V-VCE 25 |
| Motors comply with DIN EN 60034, protection class IP 55 and insulation class F. p.19 ↗ | Verify motor data plate against mains network; use appropriate motor protection switch. | V-VCA 25 and V-VCE 25 |
| Unless otherwise stated on the motor data plate, permitted tolerances are ±5 % voltage deviation and ±2 % frequency deviation. p.19 ↗ | Check supply quality before commissioning; wrong voltage/frequency can cause loss of power or damage. | V-VCA 25 and V-VCE 25 |
| Looking at the motor fan, the motor must rotate clockwise; check with a phase sequence indicator (anti-clockwise rotating field). p.21 ↗ | Verify rotation direction before continuous operation; reverse rotation can draw in media and damage the machine. | V-VCA 25 and V-VCE 25 |
| After first start, stop the motor after 2 minutes and top oil up to the upper edge of the sight glass; repeat until all oil pipes are filled. p.21 ↗ | Ensures correct oil level after initial filling of oil lines; filling point must not be open while pump is running. | V-VCA 25 and V-VCE 25 |
| Oil viscosity must comply with ISO VG 100 as per DIN ISO 3448. p.26 ↗ | Use only suitable oil; incorrect viscosity can cause overheating, insufficient pumping capacity or motor protection trips. | V-VCA 25 and V-VCE 25 |
| Check oil level every day. p.22 ↗ | Low oil level can damage the vacuum pump; investigate leaks or visible oil mist in exhaust air immediately. | V-VCA 25 and V-VCE 25 |
| Storage temperature range is -10 °C to +60 °C with max. 80 % relative air humidity; store dust-free, dry, vibration-free and protected from sun radiation. p.13 ↗ | Close openings air-tight; do not store for more than 6 months; for long-term storage over 3 months use preservation oil. | V-VCA 25 and V-VCE 25 |
| An output reduction is noticeable when installed at more than 1000 m above sea level. p.17 ↗ | Contact manufacturer for high-altitude installations. | V-VCA 25 and V-VCE 25 |
| For VCE 25 (09), fine vacuum or rough vacuum operating range is selected via an adjustment bolt in the pipe under the filter housing. p.21 ↗ | Set the adjustment bolt to the desired operating range with a screwdriver before commissioning. | V-VCE 25 (09) |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Machine is switched off by the motor protection switch p.32 ↗ | Mains voltage/ Frequency does not correspond with the motor data | Check by qualified electrician | Chapter 5.5 |
| Machine is switched off by the motor protection switch p.32 ↗ | Connection to motor terminal board is not correct | Check by qualified electrician | Chapter 5.5 |
| Machine is switched off by the motor protection switch p.32 ↗ | Motor protection switch is not set correctly | Check by qualified electrician | Chapter 5.5 |
| Machine is switched off by the motor protection switch p.32 ↗ | Motor protection switch is triggered too quickly | Use a motor protection switch with an overload-dependent delayed switch off that takes into consideration the short term excess current at start up (version with short circuit and overload trigger as per IEC 60947-4-1) | |
| Machine is switched off by the motor protection switch p.32 ↗ | Vacuum pump or its oil is too cold | Note the ambient temperature and the inlet temperature | Chapter 2.4 |
| Machine is switched off by the motor protection switch p.32 ↗ | The lubricating oil is too viscous | The oil viscosity must comply with ISO VG 100 as per DIN ISO 3448 | Chapter 7.6 |
| Machine is switched off by the motor protection switch p.32 ↗ | The oil separator insert is dirty. | Change the oil separator insert | Chapter 7.7 |
| Machine is switched off by the motor protection switch p.32 ↗ | The back pressure in the exhaust line is too high | Check the hose and/or the pipe | Chapter 5.3 |
| Pumping capacity is insufficient p.32 ↗ | The inlet pipe is too long or too narrow | Check the hose and/or the pipe | Chapter 5.3 |
| Pumping capacity is insufficient p.32 ↗ | Leak on the inlet side of the vacuum pump or in the system | Check the pipework and screw connections for leaks and check for tight fit | Chapter 7.2 |
| Pumping capacity is insufficient p.32 ↗ | The intake filter is dirty | Clean or replace the intake filter | Chapter 7.5 |
| Final pressure (max. vacuum) is not reached p.33 ↗ | Leak on the inlet side of the vacuum pump or in the system | Check the pipework and screw connections for leaks and check for tight fit | Chapter 7.2 |
| Final pressure (max. vacuum) is not reached p.33 ↗ | Incorrect oil viscosity | The oil viscosity must comply with ISO VG 100 as per DIN ISO 3448 | Chapter 7.6 |
| Machine gets too hot p.33 ↗ | Ambient or inlet temperatures too high | Ensure proper use | Chapter 2.4 |
| Machine gets too hot p.33 ↗ | Cooling air supply is obstructed | Check ambient conditions | Chapter 5.1 |
| Machine gets too hot p.33 ↗ | Cooling air supply is obstructed | Clean ventilation slots | Chapter 7.4 |
| Machine gets too hot p.33 ↗ | The lubricating oil is too viscous | The oil viscosity must comply with ISO VG 100 as per DIN ISO 3448 | Chapter 7.6 |
| Machine gets too hot p.33 ↗ | The oil separator insert is dirty. | Change the oil separator insert | Chapter 7.7 |
| Machine gets too hot p.33 ↗ | The back pressure in the exhaust line is too high | Check the hose and/or the pipe | Chapter 5.3 |
| Exhaust air contains visible oil mist p.33 ↗ | The oil separator insert is not inserted correctly. | Check that it is correctly seated | Chapter 7.7 |
| Exhaust air contains visible oil mist p.33 ↗ | Unsuitable oil is being used | Use suitable types | Chapter 7.6 |
| Exhaust air contains visible oil mist p.33 ↗ | The oil separator insert is dirty | Change the oil separator insert | Chapter 7.7 |
| Exhaust air contains visible oil mist p.33 ↗ | Ambient or inlet temperatures too high | Ensure proper use | Chapter 2.4 |
| Exhaust air contains visible oil mist p.33 ↗ | Cooling air supply is obstructed | Check ambient conditions | Chapter 5.1 |
| Exhaust air contains visible oil mist p.33 ↗ | Cooling air supply is obstructed | Clean ventilation slots | Chapter 7.2 |
| The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.33 ↗ | The coupling rubbers worn out | Replace the coupling rubbers | Chapter 7.8.2 |
| The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.33 ↗ | The pump housing is worn (chatter marks) | Repair by manufacturer or authorised workshop | Elmo Rietschle Service |
| The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.33 ↗ | The vacuum adjustment valve (if available) is vibrating | Replace the valve | Chapter 7.10 |
| The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.33 ↗ | The blades are damaged | Repair by manufacturer or authorised workshop | Elmo Rietschle Service |
| The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.33 ↗ | Vacuum pump or its oil is too cold | Note the ambient temperature and the inlet temperature | Chapter 2.4 |
| The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.33 ↗ | The lubricating oil is too viscous | The oil viscosity must comply with ISO VG 100 as per DIN ISO 3448 | Chapter 7.6 |
| Water in lubricating oil p.34 ↗ | Pump sucks in water | Install water interceptor upstream of the pump | — |
| Water in lubricating oil p.34 ↗ | The pump sucks in more water vapour than is suitable for its water vapour compatibility | Contact the manufacturer for increased gas ballast | Elmo Rietschle Service |
| Water in lubricating oil p.34 ↗ | Pump only works for a short time and therefore does not reach its normal operating temperature | Let the pump continue to run with a closed suction side after extracting the water vapour until the water has evaporated from the oil | — |