High Throughput Diffusion Pumps User Manual

Technical Reference Summary

Covered Models & Document Scope

Covers Agilent large high-throughput diffusion pumps HS-16, HS-20, HS-32 and NHS-35, including their jet assemblies, cold caps, heaters, thermal switches, cooling-water and Quick Cool coil connections, electrical wiring configurations (Delta/Wye, various source voltages), and maintenance/troubleshooting procedures.

Section-by-Section Technical Breakdown

pp. 24
Pump Operation

Describes the working principle of the diffusion pump: pump fluid is boiled, vapor travels upward through the jet assembly and is accelerated downward through jet nozzles, condensing on cooled outer walls. The vapor captures gas molecules and carries them to the ejector and out via the foreline. Large pumps use a five-stage jet assembly (four diffusion stages and one ejector stage) with water-cooled cold cap and body; optional stainless steel Quick Cool coils require an independently valved water supply.

  • Pump fluid is heated to boiling point
  • Five-stage jet assembly: four diffusion stages and one ejector stage
  • Cold cap and body are water cooled
  • Optional Quick Cool coils require independently valved water supply
pp. 24-26
Technical Specifications

Technical characteristics of the large diffusion pumps are given in Table 4, with data referring to pumps charged with DC-704 diffusion pump fluid. Air speed and throughput graphs (Figures 3–6) show performance as a function of inlet pressure, and inlet flange dimensions are given in Tables 6–12.

  • Table 4 lists technical specifications for HS-16, HS-20, HS-32, NHS-35
  • Data refers to pumps charged with DC-704 diffusion pump fluid
  • Air speed and throughput curves in Figures 3–6
  • Flange dimensions in Tables 6–12
pp. 27-28
Pump Air Speed and Throughput

Presents graphs relating inlet pressure to air speed and throughput for the large vacuum pumps (Figures 3 through 6). Figure 3 covers HS-16 at 8.1 kW; Figures 4–6 cover HS-20, HS-32 and NHS-35.

  • Figure 3: HS-16 Speed and Throughput Curves, 8.1 kW
  • Figure 4: HS-20 Speed and Throughput Curves
  • Figure 5: HS-32 Speed and Throughput Curves
  • Figure 6: NHS-35 Speed and Throughput Curves
pp. 29-32
Physical Specifications

Provides outline drawings and dimension/weight tables for each pump model with ASA flanges. Tables 5, 7, 9 and 11 give dimensions and weights for HS-16, HS-20, HS-32 and NHS-35 respectively; Tables 6, 8, 10 and 12 give flange dimensions. Connection notes indicate 3/8" FPT / G1/2 (optional) for body and foreline and 3/8" FPT for Quick Cool coils.

  • HS-16 outline and dimensions in Figure 7, Tables 5 and 6
  • HS-20 outline and dimensions in Figure 8, Tables 7 and 8
  • HS-32 outline and dimensions in Figure 9, Tables 9 and 10
  • NHS-35 outline and dimensions in Figure 10, Tables 11 and 12
  • Body and foreline connections: 3/8" FPT / G1/2 (optional); Quick Cool coils: 3/8" FPT
pp. 33
Setup and Assembly

Covers unpacking, visual inspection, and assembly steps: remove flange covers and protective plugs without damaging O-ring grooves, remove all antirust paper, verify the jet assembly is concentric and firmly seated with the ejector nozzle opposite the foreline, and clean the pump if the expected vacuum is below 10^-7 Torr.

  • Check weight in Table 4 before lifting; use power-assisted equipment
  • Remove all antirust paper from internal pump and foreline area
  • Jet assembly should be concentric, firmly seated, and held by an indexing pin
  • Clean pump if expected vacuum level is below 10^-7 Torr
pp. 34
Cleaning a New Pump

A new pump requires cleaning only if the desired vacuum is below 1x10^-7 Torr. Cleaning involves acetone and alcohol, which are toxic and explosive; safety precautions include avoiding high-temperature sources, using a well-ventilated room with a blower, and wearing protective clothing and possibly self-contained breathing apparatus.

  • Clean only if desired vacuum is below 1x10^-7 Torr
  • Acetone and alcohol are toxic and explosive
  • Use large, well-ventilated room and ventilate with a blower
  • Wear eyeshields, gloves, and other protective clothing
pp. 35
Disassembly and Reassembly for Initial Cleaning

Describes cleaning of the jet assembly, drain plugs, sight glass and pump interior. Disassembly involves removing the cold cap, jet system, fill/drain plugs and sight glass; O-rings must not be cleaned with strong solvents. Reassembly reverses the steps and requires cleaning flange and O-ring grooves with acetone or alcohol followed by drying with clean, dry, oil-free compressed air.

  • Do not use solvents on O-rings
  • Clean components with acetone followed by alcohol rinse
  • Dry all components with clean, dry, oil-free compressed air
  • Re-install O-ring gasket after cleaning flanges and grooves
pp. 36-38
System and Utility Connections

Covers vacuum connections, cooling water, and Quick Cool coil connection. The pump body must be installed vertical and plumb with the mating flange horizontal within ±1°. Cooling water must maintain inlet temperature 60–80 °F (16–27 °C) and exit water not exceeding 130 °F (54 °C). The Quick Cool coil must be connected to an open drain controlled by a separate 3-way valve.

  • Pump body vertical and plumb; mating flange horizontal ±1°
  • Cooling water inlet temperature 60 to 80 °F (16 to 27 °C)
  • Exit water temperature must not exceed 130 °F (54 °C)
  • Quick Cool coil feed controlled by separate 3-way valve (open, closed, vented)
  • Drain must be below boiler level for complete draining
pp. 39
Obtaining High Vacuum on the NHS-35

At pressures below 1x10^-6 Torr, ultimate pressure can be lowered by bypassing the cooling coils at the bend of the foreline, raising foreline temperature and providing additional degassing of fluid returning to the boiler. This configuration decreases maximum throughput capacity and should only be used if the pump is not operated near the high-pressure end of its range.

  • Applicable below 1x10^-6 Torr
  • Bypass cooling coils at foreline bend
  • Decreases maximum throughput capacity
  • Use only if pump is not operated near high-pressure end of range
pp. 39-41
Electrical Connections

Terminal connections differ by source voltage and heater configuration (Y or Delta). Table 13 lists wiring diagram locations for each pump and source voltage. Wiring steps include checking heater supply voltage, measuring branch resistance for load balance, connecting earth ground to the aluminum set screw ground lug, and wiring thermal switches to shut off power if a switch opens.

  • Do not operate pump at more than 5% over rated voltage
  • Heater resistances given on respective wiring diagrams
  • Connect earth ground wire directly to aluminum set screw ground lug
  • Wire thermal switches to control mechanism to shut off power if switch opens
  • Input wires to terminal block positions L1, L2, L3 (& N)
pp. 42
Overheating: Detection by Thermal Switches

Two normally closed thermal cutout switches monitor boiler temperature and water temperature; they are factory-set and automatically reset when temperature lowers. A rise in pump temperature can be caused by failure of cooling water flow, high inlet pressure, or low fluid level in the boiler. Thermal switches set for diffusion pump fluid are not designed to protect against overheating of backstreamed mechanical pump fluids.

  • Two normally closed thermal cutout switches: boiler and water
  • Switches are factory-set and do not require adjustment
  • Automatic reset when temperature lowers
  • Causes: cooling water failure, high inlet pressure, low fluid level
  • Not designed to protect against backstreamed mechanical pump fluids
pp. 53
Initial Vacuum Test

Section referenced in the table of contents for performing an initial vacuum test after installation.

pp. 54-56
Adding or Changing Pump Fluid

Section referenced in the table of contents covering information on fluid and diffusion pump performance, including adding or changing pump fluid.

pp. 56-58
Operation

Covers startup procedure, humid environments, and shutdown procedure. Startup procedure is referenced on page 57; humid environments on page 57; shutdown procedure on page 58.

  • Startup procedure on page 57
  • Humid environments on page 57
  • Shutdown procedure on page 58
pp. 59-60
Maintenance

Covers periodic inspections and cleaning, including cleaning safety. Periodic inspections are on page 59; cleaning and cleaning safety on page 60.

  • Periodic inspections on page 59
  • Cleaning on page 60
  • Cleaning safety on page 60
pp. 76-78
Troubleshooting

Covers leakage, outgassing, and poor pump or system performance. Leakage checks include inlet and foreline connections, drain and fill plugs, compression fittings, and threaded connections. Outgassing can be estimated by isolating the system and measuring pressure rise. Table 25 provides a troubleshooting guide with faults, probable causes, and corrective actions.

  • Check vacuum gauge performance and accuracy before step-by-step troubleshooting
  • Gas load estimated from pressure rise after isolation: p = Q/S
  • Table 25 lists faults, probable causes, and corrective actions
  • Leakage checks: inlet/foreline connections, drain/fill plugs, compression fittings, threaded connections
pp. 79
Accessories and Spare Parts

Section referenced in the table of contents listing accessories and spare parts for the high throughput diffusion pumps.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pump body must be installed vertical and plumb; mating flange on the system must be horizontal within ±1°. p.36 ↗ Correct system mount before installing the pump to avoid misalignment and vacuum leaks. HS-16, HS-20, HS-32, NHS-35
Cooling water inlet temperature must be 60 to 80 °F (16 to 27 °C); exit water temperature must not exceed 130 °F (54 °C). p.37 ↗ Recirculating water systems must provide adequate cooling and flow; use parallel connections if water pressure is low or discharge temperature routinely exceeds 130 °F. HS-16, HS-20, HS-32, NHS-35
Do not operate the pump at more than 5% over its rated voltage. p.40 ↗ Verify supply voltage and heater wiring configuration (Delta or Wye) against the appropriate wiring diagram before energizing. HS-16, HS-20, HS-32, NHS-35
Thermal switch cutout temperatures: HS-16 water 185 °F, boiler 390 °F; HS-20 water 185 °F, boiler 390 °F; HS-32 water 220 °F, boiler 550 °F; NHS-35 water 200 °F, boiler 600 °F. p.42 ↗ Wire thermal switches to shut off heater power if either switch opens; switches reset automatically when temperature lowers. HS-16, HS-20, HS-32, NHS-35
Provide an interlock to ensure heater power cannot be activated if the foreline pump is not running and/or foreline pressure is above 0.5 Torr (0.67 mbar). p.17 ↗ Prevents overpressure and explosion hazards during operation. All models
Do not run pump without cooling water, with low pump fluid level, without proper backing or holding pump, or when not evacuated below 0.5 Torr (0.67 mbar). p.19 ↗ These conditions increase the probability of explosion; ensure interlocks and operating procedures prevent them. All models
Boiler temperatures reach 530 °F (275 °C); cooling water can reach scalding temperatures. p.22 ↗ Ensure surfaces have cooled to near room temperature before touching; design water system with interlock valves so power cannot be applied unless water flows in main cooling coils. All models
Diffusion pump heaters operate at voltages up to 480 V. p.22 ↗ Design systems to prevent personnel contact with high voltages; break primary circuit before working on heater or wiring. All models
Pumps weigh in excess of 500 lbs (226.8 kg) and are 3 to 6 feet (1 to 2 meters) in largest dimension. p.22 ↗ Use power-assisted lifting equipment and trained personnel; check weight in Table 4 before lifting. All models
A new pump requires cleaning only if the desired vacuum is below 1x10^-7 Torr. p.34 ↗ Avoid unnecessary cleaning with toxic and explosive solvents if lower vacuum is acceptable. All models
Do not use solvents on O-rings; clean O-rings by wiping with lint-free cloth, washing in detergent and water, or using a small amount of pump fluid. p.35 ↗ Solvents degrade O-ring materials and reduce their ability to hold vacuum. All models
Quick Cool coil must be connected to an open drain and feed line controlled by a separate water 3-way valve (open, closed, vented to atmosphere); drain must be below boiler level. p.38 ↗ Ensures complete draining when Quick Cool water supply is shut off and pump is operating. All models with Quick Cool coil
For NHS-35 at pressures below 1x10^-6 Torr, bypassing cooling coils at the foreline bend can lower ultimate pressure but decreases maximum throughput capacity. p.39 ↗ Use this cooling configuration only if the pump is not meant to operate near the high-pressure end of its operating range. NHS-35
Heater resistances are given on respective wiring diagrams; check load balance by measuring resistance of each branch. p.41 ↗ Verify correct heater wiring and supply voltage before operation to prevent heater damage. HS-16, HS-20, HS-32, NHS-35
Connect earth ground wire directly to the aluminum set screw ground lug. p.41 ↗ Ensures electrical safety and proper bonding to earth ground system. All models
Thermal switches set to measure diffusion pump fluid temperature are not designed to protect against overheating or backstreamed mechanical pump fluids. p.42 ↗ Additional protection may be required if backstreaming of mechanical pump fluid is possible. All models
Gas load can be estimated by isolating the system from all pumps after evacuation and measuring the rate of pressure increase; p = Q/S. p.76 ↗ If gas load Q exceeds maximum throughput capability of the diffusion pump, pumping action will essentially be due to the mechanical backing pump. All models
Before step-by-step troubleshooting, check the performance and accuracy of the vacuum gauges used on the system. p.76 ↗ Avoid misdiagnosis due to inaccurate gauge readings. All models
Over time, minute leaks may develop through gaskets and may only be detectable with extremely sensitive mass spectrometer leak detectors. p.36 ↗ Use appropriate leak detection methods when troubleshooting poor system performance. All models
Tighten flange bolts evenly and compress O-ring seal until light contact is achieved between metal flanges; apply light to medium torque to fill and drain plugs and sight glass assembly. p.36 ↗ Prevents damage to sealing surfaces and ensures proper vacuum seal. All models

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Poor system pressure p.77 ↗ Leaks in system, virtual or real Locate and repair Table 25
Poor system pressure p.77 ↗ High process gas load Measure gas load, eliminate cause Table 25
Poor system pressure p.77 ↗ System dirty Clean system to reduce outgassing Table 25
Poor ultimate pressure p.77 ↗ Contaminated pump fluid Examine and clean pump; replace fluid Table 25
Poor ultimate pressure p.77 ↗ Low heat input Check voltage. Check for continuity, burned-out element, poor thermal contact Table 25
Poor ultimate pressure p.77 ↗ Inadequate cooling water flow Check water pressure. Check tubing for obstructions and backpressure Table 25
Poor ultimate pressure p.77 ↗ Cooling water is too cold Check temperature Table 25
Poor ultimate pressure p.77 ↗ Cooling water flow is too high Adjust water flow Table 25
Poor ultimate pressure p.77 ↗ High forepressure Check for leak in foreline, poor mechanical pump performance, breakdown of mechanical pump fluid Table 25
Poor ultimate pressure p.77 ↗ Water in Quick Cool coil Check and remove cause Table 25
Slow pumpdown p.77 ↗ Low heat input Check heaters Table 25
Slow pumpdown p.77 ↗ Low fluid level Add fluid Table 25
Slow pumpdown p.77 ↗ Malfunctioning pump assembly Check and repair or replace Table 25
Slow pumpdown p.77 ↗ Improperly located jets Damaged jet system Table 25
Inlet pressure surges p.77 ↗ Incorrect heater input Check and correct Table 25
Inlet pressure surges p.77 ↗ Fluid outgassing Condition fluid by operating the pump for a few hours Table 25
Inlet pressure surges p.77 ↗ Leak in system ahead of pump inlet Check and correct Table 25
High chamber contamination of the pump fluid p.78 ↗ Forepressure too high Check for leak in foreline, poor mechanical pump performance, breakdown of pump fluid, and incorrect valve operation Table 25
High chamber contamination of the pump fluid p.78 ↗ Prolonged operation in overload range Adhere to operating procedures Table 25
High chamber contamination of the pump fluid p.78 ↗ Cutting over from the backing pump too early in the pump down cycle Cut over at a lower chamber pressure Table 25
High chamber contamination of the pump fluid p.78 ↗ Improper system operation and air release procedures Adhere to operating procedures Table 25
Pump will not start p.78 ↗ Safety circuits or protective devices prevent contactor from staying closed Check utilities, flow devices switches, interlocks. Check thermostat operation Table 25

Technical Data

Parameter Value Condition
Water thermal switch cutout temperature 185 °F p.42 ↗ HS-16
Water thermal switch cutout temperature 185 °F p.42 ↗ HS-20
Water thermal switch cutout temperature 220 °F p.42 ↗ HS-32
Water thermal switch cutout temperature 200 °F p.42 ↗ NHS-35
Boiler thermal switch cutout temperature 390 °F p.42 ↗ HS-16
Boiler thermal switch cutout temperature 390 °F p.42 ↗ HS-20
Boiler thermal switch cutout temperature 550 °F p.42 ↗ HS-32
Boiler thermal switch cutout temperature 600 °F p.42 ↗ NHS-35