Introduces the CC-212B immersion circulator device
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
25...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
Pressure pump max. delivery
27 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
22 l/min
max. delivery pressure (suction)
0,4 bar
Pump connenction (optional)
M16x1 male
Bath volume
10,5 l
min. filling capacity
7 l
Width bath opening WxD/ bath depth
290x152/ 150 mm
Overall dimensions WxDxH **
350x375x375 mm
Net weight
11 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The usable temperature range depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.
Determines whether an external cooler or refrigerator is required for the intended process temperature.
Source
“Operating temperature range
25...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C”
Installation
The unit accepts 200–240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so branch protection must be sized accordingly.
Needed to plan the electrical supply and breaker rating before installation.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A”
Installation
External branch circuit protection according to UL 489 is mandatory.
Dictates the type of upstream protection device that must be installed for compliance.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10% only if they do not combine in opposite directions, e.g. -10% voltage with +3% frequency is not allowed.
Prevents specifying a supply that appears within tolerance but is actually outside the permitted combined envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order.
Avoids receiving a unit with the wrong plug for the site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The bath must be filled with at least 7 l of heat transfer fluid within its 10.5 l volume.
Sets the minimum fluid charge and expansion volume needed for safe pump operation.
Source
“Bath volume
10,5 l
min. filling capacity
7 l”
Operation
The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar.
Defines the external circuit flow and head available for sizing hoses and connected equipment.