CC-220B Immersion Circulator Datasheet

Contents

  1. Overview p. 1
    Product introduction and key features of the CC-220B circulator
  2. Spec p. 2
    Technical data according to DIN 12876 including temperature and performance
  3. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range25...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pump max. delivery27 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Bath volume17 l
min. filling capacity12 l
Width bath opening WxD/ bath depth290x329/ 150 mm
Overall dimensions WxDxH **350x555x375 mm
Net weight14 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method chosen, so the required range must be matched to water cooling or refrigerator options.

Determines whether the bath can reach sub-ambient or near-ambient temperatures for the process.

Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation

The unit requires a 200-240 V single- or two-phase supply with a minimum 10 A fuse and maximum 16 A fuse.

Electrical supply and branch protection must be sized accordingly before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

Voltage and frequency tolerances are limited to ±10% and must not combine in opposite directions.

Site power quality must be checked to avoid invalid supply conditions and equipment damage.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Safety

External branch circuit protection according to UL 489 is required.

The installer must provide a compliant branch circuit protective device.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug that must be specified at order.

The correct plug type must be ordered to match the site's electrical outlet.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Three-phase devices below 63 A are delivered with cable but without plug, and above 63 A without cable or plug.

Site wiring and termination must be planned for three-phase variants.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The bath has a minimum filling capacity of 12 l and a bath volume of 17 l.

Heat transfer fluid volume and expansion space must be planned for safe operation.

Source
“Bath volume 17 l min. filling capacity 12 l”
Operation

The pressure pump delivers up to 27 l/min at 0.7 bar and the suction pump up to 22 l/min at 0.4 bar.

External circuit pressure drop and flow requirements must be checked against pump capability.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar”
Operation

The bath bridge has a hole for a cooling probe for an immersion cooler such as TC45-TC100E.

Cooling capability can be added via an immersion cooler if the application requires it.

Source
“Bath bridge with hole for cooling probe (e.g. for immersion cooler TC45-TC100E).”
Operation

When using the Huber pump adapter, polyglycol is not permissible as heat transfer fluid.

Fluid selection must avoid polyglycol if the pump adapter is installed.

Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Operation

The ambient temperature for operation must be between 5 °C and 40 °C.

The installation location must maintain ambient conditions within this range.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Compliance

The unit is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection

The degree of protection is IP20, so the unit is not protected against water ingress.

The installation location must be dry and protected from splashing or dripping liquids.

Source
“Degree of Protection IP20”
Maintenance

A drain valve with cap is available as an accessory for emptying the bath.

Draining and fluid replacement during maintenance are facilitated by this accessory.

Source
“Drain valve with cap #6839”