Huber CS 330 OD Circulating Bath - Technical Data

Contents

  1. Overview p. 1
    Product identification and general description of the CS 330 OD unit
  2. Spec p. 2
    Technical data table according to DIN 12876 standard
  3. p. 1 table, 40 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...15 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C1,5 K
Interface digitalRS232
Safety classificationI / NFL
Kälteleistung bei Umgebungstemperatur 5°C
at 15°C40 kW
at 10°C32 kW
at 0°C22 kW
at -10°C15 kW
at -20°C10 kW
Cooling power at ambient temperature 20°C
at 15°C42 kW
at 10°C35 kW
at 0°C25 kW
at -10°C17 kW
at -20°C11 kW
Kälteleistung bei Umgebungstemperatur 35°C
at 15°C40 kW
at 10°C33 kW
at 0°C22 kW
at -10°C15 kW
at -20°C9,5 kW
Kälteleistung bei Umgebungstemperatur 40°C
at 15°C38 kW
at 10°C27 kW
at 0°C21 kW
at -10°C13 kW
at -20°C7,5 kW
Kälteleistung bei Umgebungstemperatur 50°C
at 15°C31 kW
at 10°C26 kW
at 0°C18 kW
at -10°C11 kW
at -20°C6,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity9 kg
CO2 equivalent12,57 t

Engineering Facts

12 facts · every claim linked to a page
Selection

Cooling capacity must be selected based on the actual ambient temperature at the installation site, since capacity drops significantly at higher ambient temperatures.

Prevents undersizing the chiller when installed in a warm environment.

Source
“Output data valid for: Room temperature 20°C to 50°C . If the ambient temperature rises, the cooling capacity may drop.”
Installation

The chiller is suitable for outdoor installation down to -20°C ambient, but cooling capacity is reduced above +35°C ambient.

Determines whether the unit can be placed outdoors and what capacity derating to expect.

Source
“Suitable for outside mounting (min. environmental temperature -20°C)”
Installation

Three-phase devices under 63A are supplied with cable but without a plug, so the installer must provide a suitable plug or hardwired connection.

Affects electrical installation planning and bill of materials.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are interdependent: -5% voltage with +2% frequency is not allowed, but -5% voltage with -2% frequency is allowed.

Ensures the site power quality is compatible with the chiller's EN60034-1 tolerances.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation requirements apply.

Needed for refrigerant safety assessment and compliance with local regulations.

Source
“R-449A (A1, H280)”
Compliance

The equipment emits Class A, Group 1 disturbances per EN55011, which may require EMC precautions in residential or sensitive environments.

Determines EMC compatibility and possible filtering or separation requirements on site.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.

Allows matching pump flow to the process without external throttling.

Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Operation

The remote control panel is connected via a 20 m cable, allowing the operator interface to be placed away from the chiller.

Affects control room layout and cable routing during installation.

Source
“Delivered with remote control: operating panel connected via 20m cable”
Maintenance

The recommended thermofluid is a 50:50 water-monoethylene glycol mixture.

Ensures correct fluid selection for freeze protection and heat transfer performance.

Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Installation

Space requirements around the unit must be respected according to the operating conditions referenced on the manufacturer's website.

Prevents airflow restriction and ensures adequate clearance for air-cooled condenser.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Selection

The chiller uses an air-cooled refrigeration machine with CFC- and HCFC-free refrigerant.

Confirms refrigerant compliance and that no water cooling circuit is required.

Source
“air-cooled, CFC- and HCFC-free”
Safety

The electrical components have protection classification IP54, suitable for dusty and splash-water environments.

Determines suitability for outdoor or harsh installation locations.

Source
“Protection classification of the electrical components IP54”