Product introduction and general description of the CS 50-H unit
Specp. 2
Technical data according to DIN 12876 including performance and dimensions
p. 1table, 35 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-20...85 °C
temperature set point / display
colour LED Touchscreen
Internal temperature sensor
Pt100
Temperature stability at -10°C
1,5 K
Interface digital
RS232
Safety classification
I / NFL
Heating power
2,8 kW
Cooling power at ambient temperature 20°C
.
at 15°C
8 kW
at 10°C
5 kW
at 0°C
4,2 kW
at -10°C
3,5 kW
at -20°C
2 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Refrigerant quantity
4 kg
CO2 equivalent
5,59 t
Circulation pump: max. delivery
83 l/min
max. delivery pressure
3,7 bar
Pump connection
G3/4 male
min. filling capacity
100 l
max. filling capacity
110 l
Overall dimensions WxDxH **
680x730x1520 mm
Net weight
248 kg
sound pressure level +/- 4 dB(A)
63 dB(A)
Power supply (3 Phase)
208V 3~ 60Hz
max. current (3 Phase)
14 A
Fuse (3 phase)
3x16A
power supply convertible (3 phase)
460V 3~ 60Hz
max. current convertible (3 phase)
33 A
fuse convertible (3 phase)
3x40 A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
32 °C
from Serial-No.:
1.0/20
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The chiller's cooling capacity is rated at 20°C ambient and will drop if ambient temperature rises, so size the unit with margin for the actual room temperature.
Cooling capacity must be derated for real installation ambient conditions to avoid undersizing.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit requires a 3-phase 208V 60Hz supply with 14A max current and 3x16A fusing, or a convertible 460V 3~ 60Hz configuration at 33A with 3x40A fusing.
Electrical supply, breaker and cable must be provisioned for the correct voltage/current before installation.
Source
“Power supply (3 Phase) 208V 3~ 60Hz”
Installation
Three-phase devices drawing less than 63A are delivered with cable but without a plug, so the site must provide a suitable connection.
The installer must plan for a field-supplied plug/disconnect for this 14A three-phase unit.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination must be checked (e.g. -5% voltage with +2% frequency is not allowed).
Site power quality must be verified against the allowed combined tolerance to avoid invalid supply conditions.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The chiller must be installed where ambient temperature stays between 5°C and 32°C.
Room conditions outside this range affect cooling performance and reliable operation.
Source
“min. ambient temperature 5 °C”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280 (pressurized gas), and the charge is 4 kg corresponding to 5.59 t CO2 equivalent.
Refrigerant safety class, charge and CO2 equivalent matter for handling, leak response and environmental compliance.
Source
“R-449A (A1, H280)”
Safety
The unit has IP20 protection, so it must be kept in a dry indoor location protected from water ingress.
IP20 limits installation to dry, protected areas and affects enclosure/room requirements.
Source
“Degree of Protection IP20”
Operation
The flow rate can be adjusted via the manual bypass valve on the backside of the chiller, allowing matching to the application.
Commissioning requires access to the rear bypass valve to set the circulation flow.
Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Operation
The recommended thermofluid is a 50:50 water–monoethylene glycol mixture.
Using the recommended fluid ensures freeze protection and compatibility with the -20°C operating range.
Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Performance
The circulation pump delivers up to 83 l/min at up to 3.7 bar with G3/4 male connections.
External circuit pressure drop and hose sizing must be checked against pump head and flow.
Source
“max. delivery pressure 3,7 bar”
Performance
The tank holds between 100 l and 110 l of fluid, which sets the fill volume and expansion allowance.
Fluid volume and expansion space planning depend on the filling capacity range.
Source
“min. filling capacity 100 l”
Installation
The unit is 680x730x1520 mm and weighs 248 kg net, and space requirements must be respected.
Floor loading, access routes and clearance for air-cooled condenser must be planned.
Source
“Please respect space requirements.”
Compliance
The equipment is EMC Class A Group 1, which is intended for industrial environments and may cause interference in residential areas.
Class A equipment is for industrial use; installation location and nearby sensitive electronics must be considered.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation
The unit has a digital level indicator and an atmospheric open tank, so fluid level must be monitored and the tank is not sealed.
Open tank design affects evaporation, fluid top-up and placement (no pressurization).