Product introduction and main features of KISS 212B
Specp. 2
Technical data according to DIN 12876
p. 1table, 36 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
25...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connenction (optional)
M16x1 male
Bath volume
10,5 l
min. filling capacity
5,5 l
Height of bath opening
205 mm
Width bath opening WxD/ bath depth
290x152/ 150 mm
Overall dimensions WxDxH **
350x375x375 mm
Net weight
11 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
393790
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
13 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.
Determines whether the bath can reach sub-ambient or near-ambient setpoints and what cooling utility or chiller must be specified.
Source
“Operating temperature range
25...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C”
Performance
Heating power scales with supply voltage from 1.5 kW at 200 V to 2.1 kW at 240 V, so available heating capacity depends on the actual site voltage.
Affects heat-up time and whether the unit meets process heating requirements at the customer's supply voltage.
Source
“Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW”
Installation
The unit requires a 200–240 V single- or two-phase 50/60 Hz supply with a minimum 10 A fuse and maximum 16 A fuse.
Needed to size the branch circuit, protective device, and confirm the site supply matches the unit.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A”
Installation
The enclosure is rated IP20, so it must be installed in a dry location protected from water ingress and access to live parts.
Dictates the installation environment and prohibits washdown or wet locations.
Source
“Degree of Protection
IP20”
Installation
Ambient temperature must stay between 5 °C and 40 °C for proper operation.
Room conditions outside this range can affect performance, safety, and warranty.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Selection
The bath holds 10.5 l total but requires a minimum filling of 5.5 l, and the bath opening is 290 x 152 mm with 150 mm depth.
Determines whether sample vessels and immersion coolers fit and how much thermofluid must be ordered.
Source
“Bath volume
10,5 l
min. filling capacity
5,5 l
Height of bath opening
205 mm
Width bath opening WxD/ bath depth
290x152/ 150 mm”
Selection
The pressure/suction pump delivers up to 14 l/min at 0.25 bar pressure and 10.5 l/min at 0.17 bar suction.
Used to check whether the pump can supply an external loop or jacketed vessel with adequate flow and head.
The unit is Safety Class III / FL with adjustable overtemperature protection per DIN 12876, providing fail-safe protection against overheating.
Important for risk assessment and for processes requiring fail-safe overtemperature protection.
Source
“Safety classification
III / FL”
Operation
The plastic pump is not resistant to acetone or polyglycols, so water with glycerin or ethylene glycol is recommended for freeze protection.
Prevents chemical attack and pump failure when selecting the thermofluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Compliance
Voltage tolerance is ±10% only if frequency tolerance does not run in the opposite direction; for example, -10% voltage with +3% frequency is not allowed.
Site power quality must be checked; combined out-of-tolerance voltage and frequency can damage the unit or void compliance.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
An external branch circuit protection device according to UL 489 is required.
Must be included in the electrical installation design for UL-compliant installations.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Single- and two-phase devices ship with a country-specific plug that must be specified at order; three-phase devices under 63 A ship with cable but no plug.
Avoids ordering delays and ensures the correct plug for the destination country.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The controller has an autostart function after power failure, which may restart heating and pumping automatically.
Affects process safety and restart behavior after a power interruption; may need to be considered in risk assessment.