Product introduction and main features of KISS 225B
Specp. 2
Technical data according to DIN 12876
p. 1table, 36 rows
KISS 225B Technical data according to DIN 12876
Parameter
Value
Operating temperature range
25...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connenction (optional)
M16x1 male
Bath volume
23,5 l
min. filling capacity
15 l
Height of bath opening
255 mm
Width bath opening WxD/ bath depth
290x329/ 200 mm
Overall dimensions WxDxH **
350x555x425 mm
Net weight
16 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
404517
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
14 facts · every claim linked to a page
Selection
The usable temperature range depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.
Determines whether the bath can meet the process temperature window and what cooling utility or chiller must be specified.
Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation
The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so branch protection must be sized between 10 A and 16 A.
Needed to confirm site power availability and to select the correct external fuse or circuit breaker.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation
An external branch circuit protection device according to UL 489 is mandatory.
Affects electrical panel design and compliance for the installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10% only if they do not combine in the opposite direction, so a -10% voltage with +3% frequency is not acceptable.
Prevents specifying the unit on a supply where combined voltage and frequency deviations exceed the allowed envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at order.
Ensures the correct plug is ordered and avoids rework at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic must be ordered.
Prevents premature pump failure when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than using pure water.
Guides fluid selection to avoid freezing damage when operating near or below ambient.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Selection
The bath holds 23.5 l total but requires a minimum filling of 15 l, which sets the usable working volume.
Determines whether the bath volume is sufficient for the application and how much fluid to prepare.
Source
“Bath volume 23,5 l min. filling capacity 15 l”
Performance
The pressure pump delivers up to 14 l/min at 0.25 bar, while the suction side delivers up to 10.5 l/min at 0.17 bar.
Used to check whether the pump can supply an external circuit or connected equipment.