Installation
The unit requires cooling water with a differential pressure between 3 and 6 bar, so the facility water supply must be verified before installation.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.
Source
“min. cooling water differential pressure 3 bar”
Installation
Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if condensation conditions occur.
Operating outside this range reduces cooling power and risks condensation inside the unit.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly as the setpoint decreases, from 1.2 kW at 15°C to 0.35 kW at -20°C, which must be considered when sizing the load.
The required cooling capacity at the lowest operating temperature determines whether this chiller is suitable.
Source
“at 15°C 1,2 kW”
Safety
The refrigerant is R-290 (propane, A3 flammable), so the installation area must be assessed for flammable refrigerant handling.
Flammable refrigerant classification affects site safety requirements and ventilation.
Source
“R-290 (A3, H220)”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used as heat transfer media.
Using incompatible fluids will damage the pump and void reliable operation.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Proper antifreeze selection prevents freezing and protects the cooling circuit.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The electrical supply must be 208-240 V single or two-phase at 50/60 Hz with a fuse rated between 10 A and 16 A.
Correct supply and protection sizing is required for safe electrical installation.
Source
“208-240V 1~/2~ 50/60Hz”
Compliance
Voltage and frequency tolerances must be applied simultaneously: ±5% voltage with ±2% frequency, and certain combinations are not allowed.
Non-compliant supply combinations can cause motor overheating or failure.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit has IP20 protection and must be installed in an ambient temperature between 5°C and 40°C.
IP20 offers no water protection, and ambient limits affect reliable operation.
Source
“min. ambient temperature 5 °C”
Operation
The return line temperature must not exceed 80°C even though the operating range is listed as -20...40(80)°C.
Exceeding the return line limit can damage components and is a key process constraint.
Source
“Permissible temperature in return line 80 °C”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.
Water chemistry must be compatible with these materials to avoid corrosion and fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Performance
Whisper mode reduces sound pressure to 51 dB(A) but also reduces pump delivery to 14 l/min and 0.2 bar.
Noise-sensitive installations may require whisper mode, which lowers hydraulic performance.
Source
“Delivery rate: 14 l/min”
Installation
The minimum filling capacity is 2.8 l and the expansion tank is 2.2 l, which determines the required heat transfer fluid volume.
Fluid volume and expansion capacity must be planned for correct commissioning.
Source
“min. filling capacity 2,8 l”