Introduces the Minichiller 600-H unit and its main features
Specpp. 1-2
Performance, temperature, and dimensional data per DIN 12876
p. 2table, 6 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S567846
Technical data according to DIN 12876 — Minichiller 600-H (Order-No.: 3066.0021.01)
Characteristic
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Cooling power at ambient temperature +20°C
at 15°C
0,6 kW
at 0°C
0,5 kW
at -10°C
0,35 kW
at -20°C
0,15 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,099 kg
CO2 equivalent
0,0 t
Circulation pump:
Pressure- and suction pump
max. delivery
24 l/min
max. delivery pressure
0,7 bar
max. delivery (suction)
18 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
min. filling capacity
2,8 l
Volume of expansion
2,2 l
Overall dimensions WxDxH **
280x490x424 mm
Net weight
38 kg
Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
12 A
** см. сноску в оригинале (Overall dimensions WxDxH **)
Engineering Facts
12 facts · every claim linked to a page
Selection
The unit uses R-290 (A3, H220) refrigerant with GWP 0.02 and only 0.099 kg charge, which affects safety classification and environmental compliance.
A3 flammable refrigerant requires appropriate room ventilation and ignition-source assessment during installation.
Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation
The power supply requirement is 208-240V 1~/2~ 50/60Hz with max current 12 A and a 16 A fuse, so the electrical supply must be sized accordingly.
Determines breaker rating, cable sizing and whether single- or two-phase supply is available at the site.
Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and -5% voltage with +2% frequency is explicitly not allowed.
Prevents selecting a supply point whose combined voltage/frequency deviation would exceed the permitted envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The pump connection is M16x1 male and included accessories are hose connection NW12 and sleeve nuts thread M16x1, so mating hoses must match these threads.
Ensures correct hose adapters and leak-free connections are ordered before commissioning.
Source
“Pump connection M16x1 male”
Operation
The plastic pump is not resistant to acetone and polyglycols, so water mixed with glysantine or ethylene glycol is recommended for freeze protection.
Prevents choosing a thermofluid that would chemically attack the pump and cause early failure.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
Cooling capacity is valid at room temperature 20°C and may drop if ambient temperature rises, so derating must be considered for warm rooms.
Affects whether the chiller can hold the required process temperature at the actual site ambient.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The permitted ambient temperature range is 5°C to 40°C, so the installation location must stay within these limits.
Out-of-range ambient can prevent proper cooling and may void performance claims.
Source
“min. ambient temperature 5 °C”
Compliance
The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which has implications for the electromagnetic environment of the installation.
Class A equipment may require EMC precautions in residential or light-industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The degree of protection is IP20, so the unit must be installed where it is protected from water and dust ingress.
Prevents placing the chiller in wet or dusty areas where IP20 is insufficient.
Source
“Degree of Protection IP20”
Selection
The minimum filling capacity is 2.8 l and expansion volume is 2.2 l, which sets the minimum external loop volume and expansion allowance.
Ensures the external circuit volume and expansion vessel are adequate for safe operation.
Source
“min. filling capacity 2,8 l”
Performance
The pressure-suction pump delivers max 24 l/min at 0.7 bar pressure side and 18 l/min at 0.4 bar suction side, which bounds achievable flow in the loop.
Used to check whether the pump can meet the required flow rate and pressure drop of the application.
Source
“max. delivery 24 l/min”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL, which is relevant for safe temperature limiting.
Confirms the safety cut-out can be set to protect the process and the medium from overheating.
Source
“With adjustable overtemperature protection according to DIN 12876.”