Selection
The chiller delivers only 0.4 kW of cooling at -10 °C, so low-temperature duty must be sized against this reduced capacity.
Cooling capacity drops sharply with temperature, so the process heat load at the lowest setpoint determines whether this unit is adequate.
Source
“at -10°C
0,4 kW”
Installation
The unit requires a 230 V single-phase supply with a 16 A fuse and draws up to 15.5 A, so the site circuit must be dedicated and adequately rated.
A 15.5 A draw on a 16 A fuse leaves little margin and dictates cable, breaker and socket selection during installation.
Source
“max. current
15,5 A”
Installation
Voltage and frequency tolerances must be met simultaneously, and a -5% voltage with +2% frequency combination is explicitly not allowed.
Supply quality at the installation site must be checked, since some otherwise plausible tolerance combinations are prohibited.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The chiller is air-cooled, so adequate clearance and airflow around the unit are required at the installation site.
Air-cooled condensers reject heat to the room, so insufficient space or high ambient temperature will reduce cooling capacity.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Operation
Cooling capacity is rated at 20 °C room temperature and may fall if the ambient temperature rises.
The published performance figures are only valid at 20 °C ambient, so hot rooms require derating or additional ventilation.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation
The permitted ambient temperature range is 5 to 40 °C, which bounds where the unit can be operated.
Operating above 40 °C ambient is outside specification and risks reduced cooling or equipment protection trips.
Source
“max. ambient temperature
40 °C”
Safety
The refrigerant is R-449A with GHS classification H280 and a GWP of 1397, so refrigerant handling and leak reporting obligations apply.
The refrigerant charge of 1.2 kg and its GWP affect service procedures, leak checks and environmental compliance at the site.
Source
“R-449A (A1, H280)”
Safety
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.
IP20 offers no water protection, ruling out washdown areas or damp environments during installation.
Source
“Degree of Protection
IP20”
Safety
The chiller is classified as Class A, Group 1 for electromagnetic disturbance, which affects its placement relative to sensitive equipment.
Class A equipment is intended for industrial environments and may cause interference in residential or sensitive laboratory settings.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single-phase devices are supplied with a power cable and country-specific plug, which must be specified at order time.
The plug type must be confirmed with the order, otherwise the delivered cable may not match the site socket.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The pump delivers 25 l/min maximum but only 5 l/min at 3 bar, so external circuit pressure drop must be checked against the pump curve.
A restrictive external loop will reduce flow well below the nominal 25 l/min, affecting heat transfer and temperature stability.
Source
“Delivery at 3,0 bar
5 l/min”
Maintenance
The minimum filling capacity is 2.5 l with a 2 l expansion volume, which sets the heat transfer fluid charge and expansion allowance.
Sizing the fluid charge and expansion tank correctly prevents overflow or cavitation during temperature cycling.
Source
“min. filling capacity
2,5 l”
Selection
The unit is intended for externally closed applications, which constrains the type of external circuit it may serve.
Open baths or unsealed external loops are outside the intended use and may cause fluid loss or contamination.
Source
“For externally closed applications.”