Unichiller 022w-H OLÉ Technical Data

Contents

  1. Overview p. 1
    Product name and model identification on cover page
  2. Spec p. 2
    Technical data according to DIN 12876 standard

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar, so the site water supply must be sized accordingly.

Ensures the cooling water circuit can deliver adequate flow without overpressurizing the unit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is water-cooled and must be connected to a cooling water supply with G1/2 male connections.

Determines the piping and fittings needed for installation.

Source
“Cooling water connection G1/2 male”
Operation

The cooling water inlet temperature must be between 5°C and 20°C; below the condensation point, cooling water pipes inside the unit must be insulated.

Operating outside this range can cause condensation, reduced cooling power, or increased water flow.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Affects sizing of the cooling water supply and expected performance at the site.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump specified.

Prevents pump damage and safety hazards from incompatible heat transfer fluids.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than using pure water.

Avoids freezing and damage in the cooling circuit when operating near or below 0°C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Compliance

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations (e.g., -5% voltage with +2% frequency) are not allowed.

Ensures the power supply quality meets the unit's requirements and avoids invalid operating conditions.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Prevents delays and ensures the correct plug type for the installation site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Selection

The refrigerant is R-449A with a GWP of 1397 and a charge of 0.43 kg, corresponding to 0.6 t CO2 equivalent.

Relevant for environmental compliance, refrigerant handling, and leak reporting obligations.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Operation

The unit has a minimum filling capacity of 3.8 l and an expansion tank of 1.7 l, which must be considered when filling the system.

Ensures sufficient heat transfer fluid volume and proper expansion volume for safe operation.

Source
“min. filling capacity 3,8 l”
Installation

The unit requires a 16 A fuse and has a maximum current of 15.5 A at 208-230 V, 1~/2~, 60 Hz.

Determines the electrical supply and protection sizing for installation.

Source
“max. current 15,5 A”
Operation

The pump delivers a maximum flow of 29 l/min and a maximum delivery pressure of 1 bar, with 21 l/min at 0.5 bar.

Used to assess whether the circulation pump meets the external circuit's flow and pressure requirements.

Source
“max. delivery 29 l/min”