Unichiller 022w-H Recirculating Chiller Datasheet

Contents

  1. Overview p. 1
    Product name and general description of the chiller unit
  2. Spec pp. 1-2
    Technical data, performance figures and operating parameters per DIN 12876
  3. p. 2 table, 9 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Net weight61 kg
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current15 A
Fuse16 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.2/24

Technical data according to DIN 12876 — Unichiller 022w-H (Order-No.: 3010.0122.01)

ParameterValue
Operating temperature range-10...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C at 15°C2,2 kW
Cooling power at ambient temperature +20°C at 0°C1,6 kW
Cooling power at ambient temperature +20°C at -10°C1 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,43 kg
CO2 equivalent0,6 t
UN-numberUN 2857
Circulation pumpImmersion pump
max. delivery29 l/min
max. delivery pressure1 bar
Delivery at 0,5 bar21 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C95 l/h
Consumption at water 15°C, flow 0°C96 l/h
Consumption at water 15°C, flow -10°C77 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **420x487x579 mm

** см. сноску в оригинале (на странице не раскрыта)

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water inlet temperature between 5°C and 20°C, and pipes must be insulated if water temperature can drop below the condensation point.

Defines the acceptable cooling water supply conditions and insulation requirement to prevent condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops and cooling water flow rate may increase as cooling water temperature rises, so sizing must account for actual site water temperature.

Prevents undersizing the cooling water supply and overestimating cooling capacity at elevated water temperatures.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility checks to avoid corrosion or damage.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Ensures the electrical supply stays within the allowed combined voltage and frequency envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

The plastic pump is not resistant to acetone and polyglycols, so these media must be avoided or a more resistant plastic option requested.

Prevents chemical attack on the pump when selecting heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Informs heat transfer fluid preparation for low-temperature operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Avoids delivery of the wrong plug type and ensures correct electrical connection.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Three-phase devices below 63A are supplied with cable but without plug, and above 63A without cable or plug, so site wiring must be planned accordingly.

Clarifies scope of electrical supply and need for on-site plug or wiring.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance

The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1.

Determines EMC environment suitability and potential restrictions in residential areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Space requirements must be respected according to the operating conditions referenced by the manufacturer.

Ensures adequate clearance for operation and maintenance.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Safety

The unit has adjustable overtemperature protection according to DIN 12876.

Provides a safety function that must be configured for the application.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection

The refrigeration machine is water-cooled and uses CFC- and HCFC-free refrigerant R-449A with GWP 1397.

Affects cooling water availability and environmental/regulatory considerations.

Source
“Refrigeration machine water-cooled, CFC- and HCFC-free”