Introduces the Unichiller 025T-H chiller and its main features
Specpp. 1-2
Performance, temperature, cooling capacity and dimensional data per DIN 12876
p. 2table, 4 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S479006
Technical data according to DIN 12876 — Unichiller 025T-H (Order-No.: 3054.0013.01)
Characteristic
Value
Operating temperature range
-10...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Temperature stability at -10°C
0,2 K
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
2,5 kW
Cooling power at 0°C
1,2 kW
Cooling power at -10°C
0,6 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,2 kg
CO2 equivalent
1,7 t
Circulation pump
B
max. delivery
25 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
20 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 2,0 bar
10 l/min
Delivery at 3,0 bar
5 l/min
Pump connection
G3/4 male
min. filling capacity
2,5 l
Volume of expansion
2 l
Overall dimensions WxDxH **
450x510x1230 mm
Net weight
149 kg
sound pressure level +/- 4 dB(A)
74 dB(A)
Power supply requirement
230V 1~ 50Hz
max. current
15,5 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The chiller is rated for an operating temperature range of -10 to 100 °C, so it must be matched to the process temperature window.
Defines whether the unit can cover the required process temperatures before selection.
Source
“Operating temperature range
-10...100 °C”
Performance
Cooling capacity drops sharply at low temperature (2.5 kW at 15 °C, 1.2 kW at 0 °C, 0.6 kW at -10 °C), so sizing must use the capacity at the actual operating point.
Prevents undersizing when the process runs cold rather than near ambient.
Source
“at 15°C
2,5 kW
at 0°C
1,2 kW
at -10°C
0,6 kW”
Installation
The unit requires a 230 V single-phase 50 Hz supply with up to 15.5 A and a 16 A fuse, so the electrical circuit must be sized accordingly.
Determines the dedicated circuit, breaker and cabling needed at the installation site.
Source
“Power supply requirement
230V 1~ 50Hz
max. current
15,5 A
Fuse
16 A”
Installation
Voltage and frequency tolerances are ±5% and ±2% but must not combine in the unfavorable direction, so supply quality must be verified.
Avoids operating the compressor and electronics outside their allowed supply envelope.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit is air-cooled and needs free space around it, so the installation location must respect the specified space requirements.
Insufficient clearance reduces cooling capacity and can trip the refrigeration circuit.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Operation
Cooling capacity is valid at 20 °C room temperature and may drop at higher ambient, so the ambient limit of 40 °C must be respected.
Explains why performance may be lower than the datasheet in a warm plant room.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The pump connection is G3/4 male, so hoses and adapters must match this thread.
Ensures correct fittings are ordered for the external circuit.
Source
“Pump connection
G3/4 male”
Operation
The pump delivers 25 l/min at zero head but only 5 l/min at 3 bar, so the external circuit pressure drop must be checked against the pump curve.
Prevents selecting a unit whose pump cannot overcome the external loop resistance.
Minimum filling capacity is 2.5 l with a 2 l expansion volume, so the external loop plus unit must be filled within these limits.
Avoids underfilling or overfilling the atmospheric expansion system.
Source
“min. filling capacity
2,5 l
Volume of expansion
2 l”
Safety
The refrigerant is R-449A (A1, H280) with GWP 1397 and 1.2 kg charge, so leak handling and CO2-equivalent reporting obligations apply.
Affects site safety classification and refrigerant record-keeping requirements.
Source
“Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)”
Safety
The unit is safety class III / FL with adjustable overtemperature protection per DIN 12876, so it is intended for flammable liquids with the appropriate protection setting.
Determines suitability for flammable heat transfer fluids and required safety settings.
Source
“Safety classification
III / FL”
Installation
The unit is IP20 and must be installed in a dry indoor location with ambient between 5 and 40 °C.
Rules out wet, dusty or freezing locations and defines the allowed room climate.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Compliance
The unit is EMC class A group 1, so it is intended for industrial environments and may cause interference in residential areas.
Affects installation location and possible need for filtering in sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single-phase devices are supplied with a country-specific plug that must be specified at order, so the plug type must be confirmed before delivery.
Prevents a unit arriving with a plug incompatible with the site socket.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”