Unichiller 075w-H Recirculating Chiller Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller 075w-H chiller and its main features
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 15 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -20°C240 l/h
min. cooling water differential pressure0,5 bar
max. cooling water pressure6 bar
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1050 mm
Net weight300 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)12,5 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S469065

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 075w-H (Order-No.: 3040.0010.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C
at 100°C7,5 kW
at 20°C7,5 kW
at 0°C6,1 kW
at -10°C4 kW
at -20°C2,4 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,95 kg
CO2 equivalent1,3 t
Circulation pumpE1
max. delivery48 l/min
max. delivery pressure3,4 bar
Delivery at 0,3 bar46 l/min
Delivery at 0,5 bar44 l/min
Delivery at 1,0 bar40 l/min
Delivery at 1,5 bar33 l/min
Delivery at 2,0 bar27 l/min
Delivery at 2,5 bar20 l/min
Delivery at 3,0 bar11 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C348 l/h
Consumption at water 15°C, flow 0°C330 l/h
Consumption at water 15°C, flow -10°C270 l/h

Строка «Cooling power at ambient temperature +20°C» является заголовком группы для последующих строк at 100°C / at 20°C / at 0°C / at -10°C / at -20°C; значение в её второй ячейке в исходном тексте отсутствует.

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires cooling water supply between 5°C and 20°C, so the site must provide a suitable cooling water circuit.

Determines whether the facility's cooling water supply is compatible and whether additional chilling or insulation is needed.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature falls below the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion/electrical risk when using cold cooling water.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops and cooling water flow rate may increase as cooling water temperature rises, so performance must be derated accordingly.

Affects sizing and expected heat removal capacity at the actual site cooling water conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Installation

Cooling water pressure must be between 0.5 bar minimum differential and 6 bar maximum.

Ensures adequate flow without overpressurizing the cooling water circuit.

Source
“min. cooling water differential pressure 0,5 bar max. cooling water pressure 6 bar”
Installation

The unit requires a 400V 3-phase 50Hz supply with max current 12.5 A and 3x16A fusing.

Determines electrical supply, cable and protection sizing for installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz max. current (3 Phase) 12,5 A Fuse (3 phase) 3x16A”
Installation

Three-phase devices with current consumption less than 63A are delivered with cable but without plug.

Installer must provide a suitable plug or hardwired connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the electrical supply quality is within the unit's allowed operating envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The unit has adjustable overtemperature protection according to DIN 12876.

Important for safe operation and setting appropriate protection limits for the application.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The unit is intended for externally closed applications.

Confirms the hydraulic configuration and prevents misuse with open baths.

Source
“For externally closed applications.”
Operation

The refrigerant is R-449A with GWP 1397 and CO2 equivalent 1.3 t, which matters for environmental compliance and leak management.

Relevant for refrigerant handling, leak checks and environmental reporting.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280) Global Warming Potential (GWP) 1397 Refrigerant quantity 0,95 kg CO2 equivalent 1,3 t”
Installation

The unit has IP20 protection and must be installed in an environment between 5°C and 40°C ambient temperature.

Defines environmental limits and protection against ingress for the installation location.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Overall dimensions are 740x1160x1050 mm and net weight is 300 kg, so space and floor loading must be planned.

Needed for site layout, access and structural considerations.

Source
“Overall dimensions WxDxH ** 740x1160x1050 mm Net weight 300 kg”
Compliance

The equipment is classified as EN55011 Class A Group 1 for electromagnetic disturbance.

Class A equipment may require EMC precautions in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”