Introduces the Unichiller 260Tw chiller model and its main features
Specp. 2
Technical data per DIN 12876 including performance and dimensions
p. 2table, 9 rows
p. 1table, 35 rows
Technical data according to DIN 12876
Parameter
Value
Net weight
375 kg
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
27,5 A
Fuse (3 phase)
3x32A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
S566049
Unichiller 260Tw — Technical data according to DIN 12876
Parameter
Value
Order-No.
3058.0005.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 15°C
26 kW
Cooling power at 0°C
26 kW
Cooling power at -10°C
18 kW
Cooling power at -20°C
12 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
2,55 kg
CO2 equivalent
3,6 t
Circulation pump
D3
max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 1,1 bar
202 l/min
Delivery at 1,5 bar
193 l/min
Delivery at 2,0 bar
179 l/min
Delivery at 2,5 bar
159 l/min
Delivery at 3,0 bar
140 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
77 l/min
Delivery at 4,5 bar
31 l/min
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
15 l
Volume of expansion
22 l
Overall dimensions WxDxH **
760x800x1615 mm
** см. сноску в оригинале (на странице не раскрыта)
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar at the connection.
Ensures the water circuit can deliver adequate flow without overpressurizing the unit.
Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation
The unit is water-cooled and must be supplied with cooling water between 5°C and 20°C; lower temperatures require insulation of the cooling water pipes.
Prevents condensation and ensures rated cooling performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling power ratings are valid only at 20°C room temperature, 15°C cooling water inlet, and 1 bar differential pressure.
Allows correct sizing; higher water temperatures reduce cooling capacity.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and - outlet.”
Installation
The chiller requires a 400 V 3-phase 50 Hz supply with a 3x32 A fuse and draws up to 27.5 A.
Determines electrical infrastructure and protection device sizing.
Source
“Power supply (3 Phase) 400V 3~ 50Hz max. current (3 Phase) 27,5 A Fuse (3 phase) 3x32A”
Installation
Three-phase devices under 63 A are delivered with a cable but without a plug, so the customer must provide a suitable plug.
Avoids delays at commissioning due to missing plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance
Voltage and frequency tolerances must be applied simultaneously: ±5% voltage with ±2% frequency, but opposite-sign deviations are not allowed.
Prevents invalid supply conditions that could damage the compressor or electronics.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% Example -5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant R-449A has a GWP of 1397 and a charge of 2.55 kg, corresponding to 3.6 t CO2 equivalent.
Relevant for refrigerant regulations, leak detection, and environmental compliance.
Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280) Global Warming Potential (GWP) 1397 Refrigerant quantity 2,55 kg CO2 equivalent 3,6 t”
Operation
The pump delivers 210 l/min maximum and 4.7 bar maximum pressure, with flow dropping to 31 l/min at 4.5 bar.
Needed to match external circuit pressure drop and avoid insufficient flow.
Source
“max. delivery 210 l/min max. delivery pressure 4,7 bar Delivery at 4,5 bar 31 l/min”
Installation
The unit has G1 1/4 male pump connections and G3/4 male cooling water connections.
Ensures correct hose couplings and adapters are ordered.
Source
“Pump connection G1 1/4 male Cooling water connection G3/4 male”
Operation
The minimum filling capacity is 15 l and the expansion volume is 22 l.
Determines heat transfer fluid volume and expansion tank sizing.
Source
“min. filling capacity 15 l Volume of expansion 22 l”
Installation
The unit must be operated within an ambient temperature range of 5°C to 40°C and has IP20 protection.
Defines suitable installation location and environmental limits.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Compliance
The equipment emits Class A, Group 1 disturbances per EN55011, which may require additional filtering in residential areas.
Affects EMC planning and installation in sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”