Unichiller 300T-H Datasheet

Contents

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/22

Technical data according to DIN 12876 — Unichiller 300T-H (Order-No.: 3029.0046.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C30 kW
Cooling power at 0°C18 kW
Cooling power at -10°C11 kW
Cooling power at -20°C6 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
UN-numberUN 2857
Circulation pumpD3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **905x2172x1900 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryI

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 100 °C, which defines the process window it can serve.

Confirms whether the unit covers the required process temperatures before selection.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity drops sharply with lower temperature (30 kW at 15 °C down to 6 kW at -20 °C), so sizing must use the capacity at the actual setpoint.

Prevents undersizing when the process runs at low temperatures.

Source
“at 15°C 30 kW at 0°C 18 kW at -10°C 11 kW at -20°C 6 kW”
Installation

The unit requires a 460 V three-phase 60 Hz supply, which dictates the site power infrastructure.

Electrical supply must match before installation planning.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

Three-phase devices under 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation scope and materials.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance

External branch circuit protection per UL 489 is mandatory, so the installer must provide a listed breaker.

Required for code-compliant electrical installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not combine in the unfavorable direction, so supply quality must be verified.

Prevents operation outside allowed supply conditions.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

Cooling capacity is valid at 20 °C room temperature and may drop at higher ambient, so the installation room must be considered.

Ambient conditions affect achievable performance.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit is rated IP20 and must be installed in an environment between 5 and 40 °C ambient.

Defines environmental limits and protection against ingress.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The refrigerant R-449A is classified A1 with GHS H280, so refrigerant handling and ventilation precautions apply.

Safety classification affects site risk assessment and handling.

Source
“R-449A (A1, H280)”
Installation

The pump connection is G1 1/4 male, so mating hoses and couplings must match this thread.

Ensures correct hydraulic connection hardware.

Source
“Pump connection G1 1/4 male”
Selection

Minimum filling capacity is 14 l and expansion volume is 39 l, which sets the minimum loop volume and expansion allowance.

Needed to size the external circuit and expansion volume.

Source
“min. filling capacity 14 l Volume of expansion 39 l”
Installation

Overall dimensions are 905x2172x1900 mm, so space and access must be planned accordingly.

Determines footprint and clearance requirements.

Source
“Overall dimensions WxDxH ** 905x2172x1900 mm”
Compliance

The equipment is EMC Class A Group 1, which may require attention in residential-like environments.

Affects EMC compliance and installation environment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”