Unichiller P012-H OLÉ Datasheet

Contents

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data, performance values and operating limits
  3. Spec p. 2
    DIN 12876 compliant technical parameters table
  4. p. 2 table, 4 rows
  5. p. 1 table, 39 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S457670

Unichiller P012-H OLÉ — Technical data according to DIN 12876

ParameterValue
Order-No.3009.0198.98
Operating temperature range-20...100 °C
temperature set point / displaydigital
Internal temperature sensorPt100
Resolution of display0,1 K
Interface digitalUSB (Device), RS232
Interface
Temperature stability at -10°C0,2 K
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 15°C1,2 kW
Cooling power at 0°C1 kW
Cooling power at -10°C0,7 kW
Cooling power at -20°C0,25 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,38 kg
CO2 equivalent0,5 t
Circulation pumpB
at 0,5 bar21 l/min
at 1,0 bar17 l/min
at 1,5 bar11 l/min
at 2,0 bar6 l/min
max. delivery25 l/min
max. delivery pressure2,5 bar
Pump connectionG3/4 male
min. filling capacity3,8 l
expansion tank1,7 l
Overall dimensions WxDxH **420x487x579 mm
Net weight62 kg
sound pressure level +/- 4 dB(A)59 dB(A)
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current14,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is rated for an operating temperature range of -20 to 100 °C, so it must be matched to the process temperature window.

Defines whether the chiller can cover the required process temperatures at all.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity falls from 1.2 kW at 15 °C to 0.25 kW at -20 °C, so low-temperature duty needs a larger unit or lower heat load.

Cooling capacity must be checked at the actual operating temperature, not only at nominal conditions.

Source
“at 15°C 1,2 kW at 0°C 1 kW at -10°C 0,7 kW at -20°C 0,25 kW”
Installation

The refrigeration machine is air-cooled, so sufficient air flow and ambient temperature limits must be respected at the installation site.

Air-cooled units reject heat to the room and require adequate ventilation and clearance.

Source
“Refrigeration machine air-cooled, CFC- and HCFC-free”
Installation

The permissible ambient temperature is 5 to 40 °C, so the room must stay within this range for reliable operation.

Ambient temperature outside this range can reduce cooling capacity or trip the unit.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity data are valid at 20 °C room temperature and may drop if ambient temperature rises.

Selection calculations must account for actual room temperature, not just the datasheet values.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The pump connection is G3/4 male, so hoses and adapters must match this thread.

Ensures correct fitting of hoses and avoids leaks or adapters at commissioning.

Source
“Pump connection G3/4 male”
Installation

The minimum filling capacity is 3.8 l and the expansion tank is 1.7 l, so the external loop volume and fluid expansion must be considered.

Prevents underfilling and overflow or pressure issues in closed loops.

Source
“min. filling capacity 3,8 l expansion tank 1,7 l”
Operation

The pump delivers 21 l/min at 0.5 bar but only 6 l/min at 2.0 bar, so system pressure drop directly limits available flow.

Flow must be checked against the actual pressure drop of the external circuit.

Source
“at 0,5 bar 21 l/min at 1,0 bar 17 l/min at 1,5 bar 11 l/min at 2,0 bar 6 l/min”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.

Overtemperature protection is a key safety function and must be configured during commissioning.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety

The refrigerant is R-449A with GHS classification H280, so refrigerant handling and leak response must follow safety rules.

Affects safety procedures, ventilation and service access for refrigerant work.

Source
“R-449A (A1, H280)”
Compliance

The refrigerant has a GWP of 1397 and a charge of 0.38 kg, giving 0.5 t CO2 equivalent, which matters for F-gas reporting.

Determines leak-check and documentation obligations under refrigerant regulations.

Source
“Global Warming Potential (GWP) 1397 Refrigerant quantity 0,38 kg CO2 equivalent 0,5 t”
Installation

The power supply requirement is 220-240 V single/two-phase 50/60 Hz with a maximum current of 14.5 A and a 16 A fuse.

Electrical supply, cable and protection must be sized for this current and fuse rating.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz max. current 14,5 A Fuse 16 A”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and some combinations are not allowed.

Poor site power quality can invalidate warranty or cause unreliable operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

The plastic pump is not resistant to acetone and polyglycols, so these fluids must be avoided or a more resistant pump specified.

Prevents chemical attack and premature pump failure in fluid selection.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”