Contents

  1. Overview p. 1
    Introduces the Unichiller P022w-H chiller and its main features
  2. Spec pp. 1-2
    Performance, temperature, and physical data per DIN 12876
  3. p. 2 table, 16 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -10°C77 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **420x487x579 mm
Net weight65 kg
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current14 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S614804

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller P022w-H — Technical data according to DIN 12876

ParameterValue
Order-No.3010.0154.01
Operating temperature range-10...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C at 15°C2,2 kW
Cooling power at ambient temperature +20°C at 0°C1,6 kW
Cooling power at ambient temperature +20°C at -10°C1 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,19 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pumpB
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C95 l/h
Consumption at water 15°C, flow 0°C96 l/h

Cooling power values are given at ambient temperature +20°C for the stated coolant temperatures (15°C, 0°C, -10°C).

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water supply with a differential pressure between 1 and 6 bar.

The site must provide cooling water within this pressure window; too low will trip the unit, too high may damage the circuit.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

Cooling water inlet temperature must be between 5 °C and 20 °C, and pipes must be insulated if below the condensation point.

Defines the acceptable cooling water supply temperature and the need for pipe insulation to prevent condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Rated cooling and heating output is only valid at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure.

Performance figures must be corrected if site conditions differ from these reference values.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Installation

The unit requires a 220–240 V single-phase 50 Hz supply with a 16 A fuse and draws up to 14 A.

Electrical supply and protection must be sized accordingly; a 16 A circuit is mandatory.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 14 A Fuse 16 A”
Installation

Voltage and frequency tolerances are ±5 % and ±2 % respectively, but not simultaneously in the same direction.

Supply quality must be checked; combined worst-case deviations are not permitted.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The refrigerant is R-290 (propane, A3 flammable, GHS H220), so ignition sources and ventilation must be considered.

Flammable natural refrigerant affects site safety classification and installation requirements.

Source
“R-290 (A3, H220)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so alternative fluids or a more resistant pump must be used.

Prevents chemical attack and pump failure when selecting heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Guides fluid selection to avoid freezing damage at low temperatures down to -10 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The unit is IP20 and must be operated in ambient temperatures between 5 °C and 40 °C.

Defines the permissible environment and excludes wet or dusty locations.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality may be used.

Water chemistry must be compatible with these materials to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection

The pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with flow dropping to 6 l/min at 2.0 bar.

External circuit pressure drop must be checked against the pump curve to ensure adequate flow.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”
Installation

Cooling water consumption ranges from 77 to 96 l/h depending on flow temperature, requiring adequate supply capacity.

Sizing of the cooling water supply and drain must accommodate up to about 96 l/h.

Source
“Consumption at water 15°C, flow 15°C 95 l/h”
Compliance

The unit is classified as pressure equipment category Art. 4.3 PED.

Affects documentation and conformity requirements for installation in the EU.

Source
“Pressure equipment category Art. 4.3 PED”
Maintenance

The minimum filling capacity is 3.8 l and the expansion volume is 1.7 l.

Determines the minimum heat transfer fluid charge and expansion vessel sizing.

Source
“min. filling capacity 3,8 l Volume of expansion 1,7 l”