Introduces the Unichiller P100w chiller and its main features
Specpp. 1-2
Performance, temperature, and physical specifications per DIN 12876
p. 2table, 17 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Consumption at water 15°C, flow -10°C
270 l/h
Consumption at water 15°C, flow -20°C
240 l/h
min. cooling water differential pressure
0,5 bar
max. cooling water pressure
6 bar
min. filling capacity
18 l
Volume of expansion
48 l
Overall dimensions WxDxH **
740x1160x1050 mm
Net weight
311 kg
sound pressure level +/- 4 dB(A)
58 dB(A)
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
13 A
Fuse (3 phase)
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S503730
** Please respect space requirements. See operating conditions at www.huber-online.com
Unichiller P100w — Technical data according to DIN 12876
Parameter
Value
Order-No.
3059.0011.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C
at 20°C
10 kW
at 0°C
7,8 kW
at -10°C
5,3 kW
at -20°C
3,1 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,03 kg
CO2 equivalent
1,4 t
Circulation pump
E2
max. delivery
130 l/min
max. delivery pressure
5,7 bar
Delivery at 2,4 bar
116 l/min
Delivery at 2,5 bar
115 l/min
Delivery at 3,0 bar
108 l/min
Delivery at 3,5 bar
100 l/min
Delivery at 4,0 bar
92 l/min
Delivery at 4,5 bar
80 l/min
Delivery at 5,0 bar
59 l/min
Delivery at 5,5 bar
20 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
348 l/h
Consumption at water 15°C, flow 0°C
330 l/h
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a differential pressure of at least 0.5 bar and no more than 6 bar.
Ensures the cooling water circuit is sized correctly and protected from overpressure during installation.
Source
“min. cooling water differential pressure
0,5 bar
max. cooling water pressure
6 bar”
Installation
The unit must be supplied with 400 V three-phase power and protected by a 3x16 A fuse.
Determines electrical supply, cable sizing and breaker selection for the installation site.
Source
“Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
13 A
Fuse (3 phase)
3x16A”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide the plug.
Affects electrical installation planning and procurement of the mating plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.
Prevents condensation, corrosion and loss of cooling performance during operation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity is rated at 20 °C room temperature and 15 °C cooling water inlet with 0.5 bar differential pressure, so higher water temperatures reduce capacity.
Allows correct interpretation of the published cooling power figures when selecting the unit.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 0,5 bar differential pressure between cooling water inlet and
-outlet.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility checks to avoid corrosion and fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance
Voltage and frequency tolerances must be applied together: ±5 % voltage with ±2 % frequency, but not opposite-sign deviations.
Defines acceptable site power quality and prevents invalid supply combinations.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 1.03 kg corresponding to 1.4 t CO2 equivalent.
Informs refrigerant handling, leak reporting and environmental compliance obligations.
Source
“R-449A (A1, H280)”
Installation
The unit has IP20 protection and must be installed where ambient temperature stays between 5 °C and 40 °C.
Restricts installation location to dry indoor areas with adequate ventilation and climate control.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The unit requires a minimum filling capacity of 18 l and has 48 l expansion volume, and space requirements must be respected.
Ensures the external circuit volume and installation clearances are planned correctly.
Source
“min. filling capacity
18 l
Volume of expansion
48 l”
Performance
The circulation pump delivers up to 130 l/min at 5.7 bar, but flow drops sharply at high pressure, e.g. 20 l/min at 5.5 bar.
Allows checking whether the pump can overcome the pressure drop of the external application.