Unichiller P180w Datasheet

Contents

  1. Overview p. 1
    Product introduction and key features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 7 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Overall dimensions WxDxH **940x1290x1130 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/21

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller P180w — Technical data according to DIN 12876

ParameterValue
Order-No.3041.0003.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 20°C18 kW
at 0°C10 kW
at -10°C6 kW
at -20°C3,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,49 kg
CO2 equivalent2,1 t
Circulation pumpE2
max. delivery130 l/min
max. delivery pressure5,7 bar
Delivery at 2,4 bar116 l/min
Delivery at 2,5 bar115 l/min
Delivery at 3,0 bar108 l/min
Delivery at 3,5 bar100 l/min
Delivery at 4,0 bar92 l/min
Delivery at 4,5 bar80 l/min
Delivery at 5,0 bar59 l/min
Delivery at 5,5 bar20 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
min. filling capacity18 l
Volume of expansion48 l

Order-No. 3041.0003.01; cooling power values are listed under the heading 'Cooling power at ambient temperature +20°C'.

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires cooling water inlet temperature between 5°C and 20°C, so the site must provide a suitable cooling water supply.

Cooling water availability and temperature determine feasibility of installation and required utilities.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature can fall below the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion/electrical risk in the cooling water circuit.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity drops as cooling water temperature rises, and cooling water flow rate may need to increase.

Affects sizing of cooling water supply and expected performance at site conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.

Guides water quality and material compatibility to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the electrical supply meets EN60034-1 tolerances for reliable operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices with current consumption less than 63A are delivered with cable but without plug.

Installer must plan for a suitable plug or direct connection on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Three-phase devices with current consumption greater than 63A are delivered without cable and without plug.

Requires on-site power cabling and connection by qualified personnel.

Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation must be considered.

Informs safety measures for refrigerant leak and service procedures.

Source
“R-449A (A1, H280)”
Selection

The unit has a CO2 equivalent of 2.1 t based on 1.49 kg of R-449A with GWP 1397, relevant for refrigerant regulations.

Needed for environmental compliance and leak-check obligations.

Source
“CO2 equivalent 2,1 t”
Operation

The pump delivers 130 l/min maximum and 5.7 bar maximum delivery pressure, with flow dropping at higher pressure.

Determines whether the pump can overcome system pressure drop in external closed loops.

Source
“max. delivery 130 l/min”
Installation

Pump connection is G1 1/4 male and cooling water connection is G3/4 male, so matching hoses and fittings are required.

Ensures correct plumbing adapters are available before installation.

Source
“Pump connection G1 1/4 male”
Selection

Minimum filling capacity is 18 l and expansion volume is 48 l, which must be considered for external loop volume.

Affects heat transfer fluid quantity and expansion tank sizing.

Source
“min. filling capacity 18 l”
Installation

The unit requires a 400V 3~ 50Hz three-phase power supply and has IP20 protection, so it must be installed in a dry indoor location.

Defines electrical supply and environmental protection requirements for siting.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Operation

Ambient temperature must be between 5°C and 40°C for operation.

Ensures the installation room stays within allowable ambient conditions.

Source
“min. ambient temperature 5 °C”