Induztr
Engineering Digest Document Viewer Related Models

Unimotive GL 60w

Contents

  1. Overview p. 1
    Product introduction and general description of the unit
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 22 rows
  4. p. 1 table, 24 rows

Technical data according to DIN 12876

ParameterValue
max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity200 mm²/s
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unimotive GL 60w (Order-No.: 5034.0001.01)

CharacteristicValue
Operating temperature range-45...95 °C
Temperature stability at -10°C0,05 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power48 kW
Cooling power with Water/ Glysantin 40/60 at 20°C56 kW
Cooling power with Water/ Glysantin 40/60 at 0°C48 kW
at -20°C34 kW
at -30°C27 kW
at -40°C18 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
UN-numberUN 2857
Circulation pump:MK pump
max. delivery201 l/min

Значения приведены дословно по тексту страницы; строки 'at -20°C', 'at -30°C', 'at -40°C' относятся к охлаждающей способности с Water/Glysantin 40/60.

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is designed for water-ethylene glycol mixtures with corrosion protection down to -45°C, so it suits automotive temperature simulation and material testing rather than general lab use.

Confirms the intended heat transfer fluid and application envelope before specifying the unit.

Source
“The temperature control systems are constructed for operation with Water-ethylene glycol mixture incl. corrosion protection (for example Glysantin®) down to -45°C.”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the site cooling water must be compatible with these materials.

Prevents corrosion or fouling of the water-cooled condenser from unsuitable site water.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The unit requires cooling water inlet between 5°C and 20°C, and pipes must be insulated if water temperature can fall below the condensation point.

Defines the cooling water supply conditions and condensation protection needed at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity is rated at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure, so actual capacity will differ if site conditions deviate.

Allows correct interpretation of the published cooling capacity during selection.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Performance

Higher cooling water temperature reduces cooling power and may increase cooling water flow rate, so capacity margins must account for warmer site water.

Prevents undersizing the cooling water supply or overestimating available cooling capacity.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

Three-phase devices below 63 A are supplied with cable but without plug, so the site must provide a suitable power connection.

Affects electrical installation planning and the need for a site-supplied plug or isolator.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance

Voltage and frequency tolerances are ±5% and ±2% respectively, but opposite-sign deviations such as -5% voltage with +2% frequency are not allowed.

Ensures the site power quality is checked against the permitted supply tolerance envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The pump connection is M38x1.5 male and the cooling water connection is G3/4 male, so matching hoses and adapters must be ordered.

Determines the piping and hose interfaces required for commissioning.

Source
“Pump connection M38x1,5 male”
Installation

Cooling water differential pressure must be between 1 bar and 4 bar, with maximum cooling water pressure 6 bar.

Defines the required cooling water supply pressure range for correct operation.

Source
“min. cooling water differential pressure 1 bar”
Safety

The unit uses R-744 refrigerant with GWP 1 and is classified A1/H280, and the pressure equipment category is II.

Informs refrigerant safety classification and environmental impact during selection and risk assessment.

Source
“R-744 (A1, H280)”
Operation

The unit has IP20 protection and must be operated between 5°C and 40°C ambient temperature.

Restricts installation location to a dry indoor environment within the ambient temperature range.

Source
“Degree of Protection IP20”
Maintenance

Minimum filling capacity is 8.9 l and the expansion tank holds 19.9 l, which sets the heat transfer fluid charge and expansion allowance.

Needed to plan fluid volume, expansion tank margin and service fill quantities.

Source
“min. filling capacity 8,9 l”