Unistat 405wl Technical Data

Contents

  1. Overview p. 1
    Product name and general introduction to the unit
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 19 rows
  4. p. 1 table, 28 rows

Technical data according to DIN 12876

ParameterValue
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C156 l/h
Consumption at water 15°C, flow -20°C48 l/h
Consumption at water 25°C, flow 0°C288 l/h
Verbrauch b. Wasser 25°C, Vorlauf -20°C120 l/h
min. cooling water differential pressure0,5 bar
max. cooling water pressure6 bar
min. filling capacity1,5 l
Filling capacity expansion tank2,8 l
Overall dimensions WxDxH **426x327x631 mm
Net weight60 kg
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current15,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S593531

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-45...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power3 kW
Cooling power with Thermooil at 250°C1,3 kW
Cooling power with Thermooil at 200°C1,3 kW
Cooling power with Thermooil at 100°C1,3 kW
Cooling power with Ethanol at 0°C1,3 kW
Cooling power with Ethanol at -20°C0,7 kW
Cooling power with Ethanol at -40°C0,15 kW
Refrigeration machineair- water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity0,15 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump: max. delivery55 l/min
Circulation pump: max. delivery pressure0,9 bar
Pump connectionM24x1,5 male
max. permissible kin. viscosity50 mm²/s

Order-No.: 1002.0063.01; Unistat 405wl

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires cooling water between 5 °C and 20 °C; if water is below dew point, insulate the cooling water pipes to prevent condensation.

Determines cooling water supply requirements and need for pipe insulation during installation.

Source
“The temperature control unit is designed for cooling water temperatures between 5 °C and 20 °C.”
Performance

Cooling capacity drops and cooling water flow rate may increase as cooling water temperature rises, so size the water supply accordingly.

Affects selection of cooling water supply capacity and expected performance at elevated water temperatures.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility to avoid corrosion or damage.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; combinations like -5% voltage and +2% frequency are not allowed.

Ensures the electrical supply stays within permissible limits for safe and reliable operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Prevents ordering delays and ensures correct plug for the installation site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Three-phase devices below 63 A come with cable but no plug, and above 63 A come without cable or plug, so site wiring must be planned.

Clarifies scope of electrical supply and need for on-site plug/connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

For external open baths the expansion tank is blocked off, making the thermostat atmospherically sealed and allowing it to be located below or above the application level.

Enables flexible placement of the unit relative to the application in open-bath setups.

Source
“For external open baths the expansion tank will be blocked off.”
Selection

The pump connection is M24x1.5 male, and pipework/flexible hoses must use ball socket per DIN 3863 and sleeve nut per DIN 3870.

Ensures correct mating fittings and hoses are selected for leak-free connection.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Safety

The refrigerant is R-1270 (A3, H220) with GWP 0 and 0.15 kg charge, classified as flammable, requiring appropriate handling and ventilation.

Informs safety measures for refrigerant handling and room ventilation during installation and service.

Source
“R-1270 (A3, H220)”
Installation

Minimum cooling water differential pressure is 0.5 bar and maximum cooling water pressure is 6 bar.

Sets the required cooling water supply pressure range for proper operation.

Source
“min. cooling water differential pressure”
Operation

The unit is rated IP20 and must operate between 5 °C and 40 °C ambient temperature.

Defines environmental limits and protection level for installation location.

Source
“min. ambient temperature”
Selection

Maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.

Guides selection of suitable heat transfer fluid to avoid pump overload.

Source
“max. permissible kin. viscosity”