Unistat 525w Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat 525w
  2. Spec pp. 1-2
    Technical performance data and operating specifications
  3. Spec p. 2
    Detailed parameters according to DIN 12876 standard
  4. p. 2 table, 20 rows
  5. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C792 l/h
Consumption at water 15°C, flow 0°C430 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity5,1 l
Filling capacity expansion tank18 l
Overall dimensions WxDxH **540x604x1332 mm
Net weight215 kg
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply requirement400V 3~ 50Hz
max. current13,5 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S393240

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C10 kW
Cooling power with Thermooil at 200°C10 kW
Cooling power with Thermooil at 100°C10 kW
Cooling power with Thermooil at 20°C10 kW
Cooling power with Ethanol at 0°C7 kW
Cooling power with Ethanol at -20°C4,2 kW
Cooling power with Ethanol at -40°C1,5 kW
Cooling power with Ethanol at -50°C0,65 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,44 kg
CO2 equivalent3,1 t
Circulation pump: MK pump max. delivery79 l/min
Circulation pump: MK pump max. delivery pressure1,5 bar
Delivery at 0,25 bar72 l/min
Delivery at 0,5 bar63 l/min
Delivery at 0,75 bar54 l/min
Delivery at 1,0 bar43 l/min
Delivery at 1,25 bar29 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and breaker rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 13,5 A Fuse 3x16A”
Installation

Cooling water must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and pipes must be insulated if water can drop below the dew point.

Defines the cooling water utility requirements and condensation risk that affect piping design.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, so higher water temperatures reduce available cooling.

Allows correct interpretation of the published cooling power figures for the actual site conditions.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Prevents corrosion or damage by selecting suitable water quality and materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Compliance

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.

Guides assessment of supply quality and whether the site power is acceptable.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the need to procure a plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863, and hoses must use ball socket per DIN 3863 and sleeve nut per DIN 3870.

Ensures correct mating fittings and hoses for leak-free connection.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The unit is rated for ambient temperatures from 5 °C to 40 °C, so the installation room must stay within this range.

Defines the allowable operating environment and ventilation/room requirements.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The refrigerant R-452A is classified A1 with GHS hazard H280, and the charge is 1.44 kg corresponding to 3.1 t CO2 equivalent.

Informs refrigerant safety classification and environmental reporting obligations.

Source
“R-452A (A1, H280)”
Selection

The unit provides 6 kW heating and 10 kW cooling with thermooil, with cooling dropping to 0.65 kW at -50 °C with ethanol.

Allows matching the unit to the required process load across the temperature range.

Source
“Cooling power with Ethanol at -50°C 0,65 kW”
Selection

The magnetic-coupled stainless steel pump delivers up to 79 l/min and 1.5 bar, with 43 l/min at 1.0 bar.

Supports checking whether the pump curve suits the external circuit pressure drop.

Source
“max. delivery 79 l/min max. delivery pressure 1,5 bar”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.

Restricts installation location and environmental protection requirements.

Source
“Degree of Protection IP20”
Maintenance

The minimum filling capacity is 5.1 l and the expansion tank holds 18 l, which sets the heat transfer fluid charge and expansion allowance.

Determines fluid volume to prepare and expansion volume available.

Source
“min. filling capacity 5,1 l Filling capacity expansion tank 18 l”
Installation

The unit weighs 215 kg and measures 540x604x1332 mm, so floor loading and access routes must be planned.

Affects handling, rigging and space requirements at the installation site.

Source
“Overall dimensions WxDxH ** 540x604x1332 mm Net weight 215 kg”