Product introduction and general description of the Unistat 527w unit
Specp. 2
Technical data and performance specifications according to DIN 12876
p. 2table, 20 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Pump connection
M38x1,5 male
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
880 l/h
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,2 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
730x804x1738 mm
Net weight
448 kg
Power supply requirement
400V 3~ 50Hz
max. current
31,5 A
Fuse
3x32A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S550886
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Unistat 527w
Order-No.
5001.0001.01
Operating temperature range
-55...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Cooling power with Thermooil at 250°C
12 kW
Cooling power with Thermooil at 200°C
12 kW
Cooling power with Thermooil at 100°C
12 kW
Cooling power with Thermooil at 20°C
12 kW
Cooling power with Ethanol at 0°C
12 kW
Cooling power with Ethanol at -20°C
6 kW
Cooling power with Ethanol at -40°C
2 kW
Cooling power with Ethanol at -50°C
1 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
2,9 kg
CO2 equivalent
6,2 t
Circulation pump: MK pump max. delivery
196 l/min
Circulation pump: MK pump max. delivery pressure
2,5 bar
Delivery at 0,5 bar
175 l/min
Delivery at 1,0 bar
152 l/min
Delivery at 1,5 bar
123 l/min
Delivery at 2,0 bar
86 l/min
Engineering Facts
13 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 3x32 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable cross-section and protection device needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz max. current 31,5 A Fuse 3x32A”
Installation
The cooling water circuit must be supplied at 5–20 °C with 1–4 bar differential pressure and no more than 6 bar static pressure.
Cooling water supply conditions directly affect achievable cooling capacity and safe operation.
Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Installation
Cooling water pipes must be insulated if the cooling water temperature can fall below the condensation point.
Prevents condensation and corrosion on the cooling water circuit at low water temperatures.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling capacity is rated at 20 °C room temperature and 15 °C cooling water inlet with 1 bar differential pressure, and drops as cooling water temperature rises.
The published 12 kW cooling capacity is only valid under the stated water conditions; higher water temperature reduces usable capacity.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The pump connection is M38x1.5 male with a Y-shaped bore (60°) per DIN 3863, requiring matching ball-socket fittings and sleeve nuts per DIN 3870.
Ensures correct mating hoses and fittings are ordered and leak-free connections are made.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Safety
Leakage current exceeds 3.5 mA, which may require a suitable residual current device or protective earthing arrangement.
Affects electrical protection design and compliance with local installation rules.
Source
“Attention: leakage current > 3,5mA”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Site power quality must be checked; some voltage/frequency combinations are outside the permitted operating envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.
The installer must plan and supply the correct three-phase plug for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit is designed for cooling water intake between 5 °C and 20 °C, so warmer water supplies are outside the intended operating range.
Cooling water above 20 °C may prevent the unit from reaching specified performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Selection
The refrigerant is R-452A with GWP 2141 and a charge of 2.9 kg, giving a CO2 equivalent of 6.2 t.
Relevant for refrigerant regulations, leak reporting thresholds and environmental compliance.
Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 2,9 kg CO2 equivalent 6,2 t”
Selection
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.
IP20 offers no water protection, and ambient limits constrain where the unit can be installed.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation
The minimum filling capacity is 8.2 l and the expansion tank holds 19.9 l, which sets the heat transfer fluid volume to prepare.
Needed to plan heat transfer fluid quantity and expansion volume for the loop.
Source
“min. filling capacity 8,2 l Filling capacity expansion tank 19,9 l”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Incompatible water chemistry can corrode or damage the cooling circuit materials.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”