Unistat 545w Technical Data

Contents

  1. Overview p. 1
    Introduces the Unistat 545w unit and its main features
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 20 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C1630 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity9,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight479 kg
Power supply requirement400V 3~ 50Hz
max. current58 A
Fuse3x63A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S565746

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicUnistat 545w
Order-No.5012.0001.01
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 250°C35 kW
Cooling power with Thermooil at 200°C35 kW
Cooling power with Thermooil at 100°C35 kW
Cooling power with Thermooil at 20°C35 kW
Cooling power with Ethanol at 0°C32 kW
Cooling power with Ethanol at -20°C16 kW
Cooling power with Ethanol at -40°C4 kW
Cooling power with Ethanol at -50°C1,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity3,6 kg
CO2 equivalent7,7 t
Circulation pump:MK pump
max. delivery196 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V 3-phase 50 Hz supply with a 3x63 A fuse, so the electrical installation must be sized for a 58 A maximum current.

Determines the site power supply, cable cross-section and upstream protection needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 58 A Fuse 3x63A”
Installation

The cooling water circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar static pressure, with 1630 l/h consumption at 15 °C.

Sizing of the cooling water supply, pump and piping must meet these pressure and flow limits.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Installation

Cooling water intake must stay between 5 °C and 20 °C, and pipes must be insulated if water temperature can fall below the dew point.

Out-of-range cooling water reduces cooling capacity and risks condensation, so the facility water supply must be checked.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity is rated at 35 kW with Thermooil from 20 °C to 250 °C but drops sharply with ethanol to 1.5 kW at -50 °C.

The heat transfer fluid choice and low-temperature process load must be matched to the actual available cooling power.

Source
“Cooling power with Ethanol at -50°C 1,5 kW”
Selection

The circulation pump delivers 196 l/min maximum but only 86 l/min at 2.0 bar, so external circuit pressure drop must be checked against the pump curve.

External loop resistance determines whether the required flow can actually be achieved.

Source
“max. delivery 196 l/min max. delivery pressure 2,5 bar Delivery at 2,0 bar 86 l/min”
Safety

The refrigerant R-452A is classified A1 with GHS hazard H280 and has a GWP of 2141 with 7.7 t CO2 equivalent, which affects refrigerant handling and reporting obligations.

Refrigerant charge, GWP and CO2 equivalent must be considered for leak detection, F-gas reporting and room ventilation.

Source
“R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 3,6 kg CO2 equivalent 7,7 t”
Installation

The unit is IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in a wet or dusty environment.

Site ambient conditions and enclosure protection dictate where the unit can be installed.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit weighs 479 kg and measures 730x804x1738 mm, so floor loading and access route must be verified.

Transport, floor strength and doorway clearances must be planned before delivery.

Source
“Overall dimensions WxDxH ** 730x804x1738 mm Net weight 479 kg”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide the final connection.

The electrical contractor must supply and install the appropriate plug or hard-wired connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.

Site power quality must be checked, since some voltage/frequency combinations invalidate the warranty and safe operation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.

Affects earth connection design and choice of residual current device in the electrical installation.

Source
“Attention: leakage current > 3,5mA”
Operation

The unit has an adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

The overtemperature cut-out must be set correctly for the application to protect the process and the operator.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Water quality and treatment must be compatible with these materials to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection

The unit is intended for both externally closed and externally open applications, which affects pump and expansion tank sizing.

Confirms suitability for open bath or external loop use, which influences installation layout and filling.

Source
“As well as for externally closed and also externally open applications.”