Induztr
Engineering Digest Document Viewer Related Models

Unistat 625w Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat 625w unit
  2. Spec p. 2
    Technical data according to DIN 12876 including performance parameters
  3. p. 2 table, 13 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply requirement460V 3~ 60Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat 625w (Order-No.: 5017.0102.01)

CharacteristicValue
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C16 kW
Cooling power with Thermooil at 100°C16 kW
Cooling power with Thermooil at 20°C16 kW
Cooling power with Ethanol at 0°C16 kW
Cooling power with Ethanol at -20°C15 kW
Cooling power with Ethanol at -40°C6,4 kW
Cooling power with Ethanol at -60°C1,7 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Circulation pump:MK pump
max. delivery196 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male

Таблица транспонирована: в исходнике характеристики — строки, значения — колонка.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a water-cooled cooling machine, so a cooling water supply with 1–4 bar differential pressure and max 6 bar pressure must be available at the installation site.

Cooling water supply pressure and differential must be verified before installation to avoid insufficient cooling or overpressure damage.

Source
“min. cooling water differential pressure 1 bar”
Installation

The cooling water intake temperature must be between 5°C and 20°C, and pipes must be insulated if condensation conditions occur.

Site cooling water temperature directly affects achievable cooling power and condensation risk.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly at low temperatures with ethanol, from 15 kW at -20°C to 1.7 kW at -60°C, so low-temperature process loads must be sized accordingly.

Engineers must check that the required heat removal at the lowest process temperature is within the reduced cooling capacity.

Source
“at -60°C 1,7 kW”
Installation

The unit requires a 460V 3~ 60Hz power supply, which must be confirmed at the site before ordering and installation.

Mismatch of voltage, phase, or frequency will prevent commissioning and may damage the equipment.

Source
“Power supply requirement 460V 3~ 60Hz”
Compliance

External branch circuit protection according to UL 489 is required, so the electrical installation must include a suitable UL-listed breaker.

This is a mandatory electrical compliance requirement for installation in the target market.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Supply quality must be checked because some voltage/frequency combinations are explicitly prohibited.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices with current consumption less than 63A are delivered with cable but without plug, so the customer must provide a suitable plug and connection.

Electrical installation planning must include procurement and wiring of the plug for this unit.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, which limits fluid selection at low temperatures.

Choosing a fluid with too high viscosity at operating temperature can overload the pump and reduce circulation.

Source
“max. permissible kin. viscosity 100 mm²/s”
Installation

The pump connection is M38x1.5 male and the cooling water connection is G3/4 male, so matching hoses and adapters must be prepared.

Connection thread compatibility must be confirmed before installation to avoid delays and leaks.

Source
“Pump connection M38x1,5 male”
Safety

The unit has safety classification III / FL according to DIN 12876 and adjustable overtemperature protection, which is relevant for flammable or sensitive applications.

Safety class determines suitability for use with flammable fluids and required protective measures.

Source
“Safety classification III / FL”
Operation

The refrigerant R-452A has a GWP of 2141, which affects environmental compliance and potential refrigerant handling obligations.

High GWP refrigerant may trigger regulatory, reporting, or leakage-control requirements at the site.

Source
“Global Warming Potential (GWP) 2141”
Installation

The unit must be operated within an ambient temperature range of 5°C to 40°C, so the installation location must be controlled accordingly.

Ambient conditions outside this range can affect performance, reliability, and warranty.

Source
“min. ambient temperature 5 °C”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only compatible cooling water may be used.

Water quality must be checked to prevent corrosion and fouling of these materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The unit requires minimum filling capacity of 8.9 l and has an expansion tank capacity of 19.9 l, which must be considered when preparing the fluid circuit.

Insufficient fluid volume can cause pump cavitation and poor temperature control.

Source
“min. filling capacity 8,9 l”