Unistat GL P530w Technical Data Sheet

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat GL P530w unit
  2. Spec pp. 1-2
    Technical data, performance figures and operating parameters per DIN 12876
  3. p. 2 table, 19 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C620 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat GL P530w — Technical data according to DIN 12876

CharacteristicValue
Order-No.5025.0002.01
Operating temperature range-50...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 200°C12 kW
Cooling power with Thermooil at 100°C12 kW
Cooling power with Thermooil at 20°C12 kW
Cooling power with Ethanol at 0°C10 kW
Cooling power with Ethanol at -20°C7 kW
Cooling power with Ethanol at -40°C2,5 kW
Cooling power with Ethanol at -50°C1,5 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
Circulation pumpMK pump
max. delivery201 l/min
max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase power connection.

Determines electrical infrastructure and whether a plug is supplied (three-phase <63 A comes with cable but no plug).

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Cooling water must be supplied at 5–20 °C with 1–4 bar differential pressure and max 6 bar pressure, so the facility water system must meet these limits.

Insufficient or out-of-range cooling water will reduce cooling capacity or prevent safe operation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water pipes must be insulated if water temperature can fall below the dew point to prevent condensation.

Prevents condensation damage and electrical/safety issues in the installation.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.

Guides water quality and material compatibility to avoid corrosion or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The pump connection is M38x1.5 male and pipework must use ball socket per DIN 3863 and sleeve nut per DIN 3870.

Ensures correct fittings and leak-free connection of external circuits.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Installation

The unit needs space per operating conditions and has overall dimensions 730x804x1738 mm, so layout must accommodate it.

Prevents installation in insufficient space and supports service access.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Operation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Defines acceptable supply quality; out-of-tolerance supply can damage the unit or void performance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

Ambient temperature must be between 5 °C and 40 °C for proper operation.

Operating outside this range may reduce performance or cause alarms.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity depends strongly on coolant and temperature, e.g. 12 kW with Thermooil at 20–200 °C but only 1.5 kW with ethanol at -50 °C.

Critical for selecting the unit for low-temperature process loads.

Source
“Cooling power with Ethanol at -50°C 1,5 kW”
Performance

Pump delivery drops with pressure: 201 l/min max, 179 l/min at 1.0 bar, 87 l/min at 4.0 bar.

Needed to match external circuit pressure drop and required flow.

Source
“Delivery at 4,0 bar 87 l/min”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

Supports safe operation and compliance with laboratory safety requirements.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety

Refrigerant is R-744 (A1, H280) with GWP 1, which affects handling and environmental compliance.

Relevant for safety classification, ventilation and environmental reporting.

Source
“R-744 (A1, H280)”
Compliance

EMC classification is Class A, Group 1 per EN55011, which may impose installation restrictions in residential environments.

Class A equipment may not be suitable for residential use without additional measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

Cooling capacity drops and cooling water flow may need to increase as cooling water temperature rises.

Affects sizing of cooling water supply and expected performance drift.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”