Introduces the Unistat GL P550w unit and its main features
Specpp. 1-2
Performance, temperature range, and technical data per DIN 12876
Specp. 2
Physical dimensions, connections, and installation data
p. 2table, 23 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 3,0 bar
124 l/min
Delivery at 4,0 bar
87 l/min
Delivery at 5,0 bar
42 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
3100 l/h
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
918x963x1771 mm
Net weight
799 kg
Power supply requirement
400V 3~ 50Hz
max. current
61,5 A
Fuse
3x63A
Pressure equipment category
II
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S573295
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unistat GL P550w (Order-No.: 5023.0002.01)
Characteristic
Value
Operating temperature range
-50...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
24 kW
Cooling power with Thermooil at 200°C
41 kW
Cooling power with Thermooil at 100°C
41 kW
Cooling power with Thermooil at 20°C
41 kW
Cooling power with Ethanol at 0°C
37 kW
Cooling power with Ethanol at -20°C
22 kW
Cooling power with Ethanol at -40°C
10 kW
Cooling power with Ethanol at -50°C
6 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-744 (A1, H280)
Global Warming Potential (GWP)
1
Refrigerant quantity
8,7 kg
CO2 equivalent
0,0 t
Circulation pump: MK pump
max. delivery
201 l/min
max. delivery pressure
5,3 bar
Delivery at 1,0 bar
179 l/min
Delivery at 2,0 bar
154 l/min
Значения приведены дословно по тексту страницы; строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' в оригинале напечатаны как группирующие заголовки с подстроками температур.
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V 3-phase 50 Hz supply with a 3x63 A fuse, so the electrical installation must be sized for a 61.5 A maximum current.
Determines the site power distribution, cable cross-section and breaker rating needed before delivery.
Source
“Power supply requirement 400V 3~ 50Hz max. current 61,5 A Fuse 3x63A”
Installation
Cooling water must be supplied at 5–20 °C with 1–4 bar differential pressure and no more than 6 bar static pressure, so the site water system must be checked against these limits.
Insufficient or excessive cooling water pressure will prevent rated cooling performance or damage the water circuit.
Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used to avoid corrosion.
Guides water quality and material compatibility decisions for the plant cooling water supply.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Installation
If cooling water temperature can fall below the condensation point, the cooling water pipes inside the unit must be insulated.
Prevents condensation and corrosion inside the unit when cold cooling water is used.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the site plug.
Affects electrical installation planning and the parts needed on site before commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for this higher leakage.
Impacts the choice of residual current device and earthing arrangement for safe operation.
Source
“Attention: leakage current > 3,5mA”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
The site power quality must be verified against this combined tolerance before installation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance
The equipment is EMC classified as Class A, Group 1 to EN55011, which is intended for industrial environments.
Class A equipment may not be suitable for residential or light-industrial locations without additional measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection
The pump delivers 201 l/min at zero head but only 42 l/min at 5.0 bar, so the external circuit pressure drop must be matched to the pump curve.
Ensures the selected model can actually achieve the required flow through a high-pressure-drop external loop.
Source
“Delivery at 5,0 bar 42 l/min”
Selection
The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, which limits fluid choice at low temperatures.
Prevents pump overload and poor circulation when selecting a cold heat transfer fluid.
Source
“max. permissible kin. viscosity 100 mm²/s”
Installation
The unit needs at least 8.9 l filling capacity and has a 19.9 l expansion tank, so the external loop volume must be planned accordingly.
Determines the minimum fluid charge and expansion volume for the installed system.
Source
“min. filling capacity 8,9 l Filling capacity expansion tank 19,9 l”
Installation
The unit weighs 799 kg and measures 918x963x1771 mm, so floor loading and access routes must be verified.
Needed for transport planning, doorway clearance and floor load assessment.
Source
“Overall dimensions WxDxH ** 918x963x1771 mm Net weight 799 kg”
Operation
The unit is rated for ambient temperatures from 5 °C to 40 °C, so the installation room must stay within this range.
Operating outside this range may reduce performance or trip the unit.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection
The unit uses R-744 (CO2) with GWP 1 and 8.7 kg charge, giving a CO2 equivalent of 0.0 t.
Relevant for refrigerant regulations, safety classification and environmental reporting.
Source
“Refrigerant (ASHRAE, GHS) R-744 (A1, H280) Global Warming Potential (GWP) 1 Refrigerant quantity 8,7 kg CO2 equivalent 0,0 t”